Supercritical fluid extraction of organics from coal gasification wastewaters using carbon dioxide
by Ambrey Gordon Gartner
A thesis submitted in partial fulfillment of the requirements for the degree of Master of Science in
Chemical Engineering
Montana State University
© Copyright by Ambrey Gordon Gartner (1985)
Abstract:
Existing chemical extraction techniques for the removal of phenols from coal gasification wastewater
require a moderately expensive solvent and additional distillation equipment for solvent recovery. This
research was designed to determine if supercritical carbon dioxide, a non-toxic, non-flammable, and
relatively inexpensive byproduct of coal gasification, could be used for phenol extraction.
Experiments were run to investigate phase equilibria associated with the two phase system consisting
of water, phenol, and carbon dioxide. System pressures varied between 1100 and 3000 psi at a
temperature of 40 degrees Celsius. Phenol concentrations in the water phase varied from 600 to 8500
mg/l. Phase equilibrium data were first used to determine the ability of supercritical carbon dioxide to
extract phenol from water, and later to determine the extraction column characteristics.
Extraction equipment was built to test the feasibility of continuous extraction of phenol from water and
wastewater. Twenty-three successful experiments were run, including one counter-current and
seventeen co-current experiments using phenol/water standards of known initial concentration, and five
co-current experiments using wastewater. The best extraction capability was obtained using
counter-current extraction techniques.
Extraction efficiencies, based on the approach to theoretical equilibrium, were determined for each of
the co-current experiments. These efficiencies ranged from 81% to 145% for the phenol/water standard
experiments, and from 76% to 163% for wastewater experiments. Relative extraction efficiencies were
based on Run No. 5, which had the highest efficiency of a phenol/water standard experiment (145%).
Extraction of phenol and phenol derivatives from coal gasification wastewater appeared to be higher
than would be expected, based on equilibrium data. This could be due to additional experimental error
involved in measuring the original phenol concentration in the wastewater. Also, the existence of
co-solvents in the wastewater could enhance the solvent power of supercritical carbon dioxide.
SUPERCRITICAL FLUID EXTRACTION OF ORGANICS FROM COAL
GASIFICATION WASTEWATERS USING CARBON DIOXIDE
by
Ambrey Gordon Gartner
A thesis submitted in partial fulfillment
of the requirements for the degree
of
Master of Science
in
Chemical Engineering
MONTANA STATE UNIVERSITY
Bozeman, Montana
October, 1985
N372om
Oo-p" oV
i i
APPROVAL
of a thesis submitted by
Ambrey Gordon Gartner
This thesis has been read by each member of the
thesis committee and has been found to be satisfactory
regarding content, English usage, format, citations,
bibliographic style, and consistency, and is ready for
submission to the college of Graduate Studies.
I o / z.3 / <0
Date Chairperson, Graduate Committee
Approved for the Major Department
Date He?ad, Major Department
Approved for the College of Graduate Studies
//- /S' .
Date Graduate Dean
STATEMENT OF PERMISSION TO USE
In presenting this thesis in partial fulfillment of
the requirements for a master's degree at Montana State
University, I agree that the Library shall make it
available to borrowers under rules of the Library. Brief
quotations from this thesis are allowable without special
permission, provided that accurate acknowledgement of
source is made.
Permission for extensive quotation from or
reproduction of this thesis may be granted by my major
professor, or in his absence, by the Director of Libraries
when, in the opinion of either, the proposed use of the
material is for scholarly purposes. Any copying or use of
the material in this thesis for financial gain shall not
be allowed without my written permission.
Signature
VACKNOWLEDGEMENTS
I would like to express my appreciation to the
Chemical Engineering Department faculty at Montana State
University for their guidance and help during the course of
my research. Special appreciation.is extended to Dr. Warren
P. Scarrah for his extra time and patience.
In addition I would to thank Lyman Fellows for his
craftsmanship and advice, Drs. Lloyd Berg and Phil
McCandless for their gas chromatograph expertise, and Dr.
Amirtharajah, Margie Reagen, and Barb Carlson for their
assistance in sample analyses and the use of Environmental
Engineering lab facilities.
Deepest appreciation is extended to my wife, Sharon,
whose support made all the work possible, and to Rachel and
Benjamin, who gave the work meaning.
vi
TABLE OF CONTENTS
PAGE
LIST OF TABLES . . .................... viii
LIST OF F I G U R E S ..................................... ix
ABSTRACT . . . . . . . xii
INTRODUCTION AND PREVIOUS RESEARCH ................ I
Introduction ...................... I
Supercritical Fluids ............................ 3
Extraction with Supercritical Fluids . . . . . . . . 7
Carbon Dioxide as a Solvent . .............. .. . 10
Previous Research............................... 14
RESEARCH OBJECTIVE ................................ 17
MATERIALS, EQUIPMENT, AND PROCEDURE . . . ........ 18
Materials............ 18
Equipment.............. '...................... 18
Equilibrium Data Equipment . . ................ 18
Extraction Apparatus .......................... 19
Analytical Equipment .......................... 27
Experimental Procedures ........ 28
Phase Equilibrium Run Procedures .............. 28
Extraction Run Preparations . ................ 29
Extraction Run Procedures .................... 31
Sample Analyses . ............................ 34
• EXPERIMENTAL RESULTS AND DATA ANALYSIS . ; ........ 38
Introduction .................................... 38
Reliability of D a t a ............................ 39
Phase Equilibria................................ 46
Counter-current Extraction ........ 54
Phenol/Water Co-current Extraction .............. 63
Wastewater Co-current Extraction . .............. 80
SUMMARY . ............................ 84
PAGE
RECOMMENDATIONS FOR FUTURE STUDY . . . . . . . w ... . 88
NOTATIONS........................................ 91
REFERENCES CITED.................... '............. 9 5
APPENDICES.......................................... 100
Appendix A. Analytical Procedures . . ............ 101
Ammonia Analysis Using the Orion 901 Analyzer . 102
Chemical Oxygen Demand (COD) - Rapid Method . . 104
Appendix B. Sample Calculations ................ 107
Calculation of Phenol Mole Fractions . . . . . . . 108
Determination of Actual Distribution
Coefficient.................................. HO
Determination of Theoretical Distribution
Coefficient.................................. Ill
'Determination of Extraction Efficiency,
Relative Efficiency, and Relative Time . . . 112
Determination of Absolute Phenol Extraction . . 113
vii
viii
LIST OF TABLES
TABLE PAGE
1. Comparison of Carbon Dioxide Fluid Properties . . 12
2. Heat of Vaporization of Carbon D i o x i d e ........ 12
3. References with Examples of Supercritical Fluid
Extraction Using Carbon Dioxide............. 13
4. References with Solubility of Solids and Liquids
in Supercritical Carbon Dioxide ................. 15
5. Raw Phase Equilibrium Data for the Phenol/Water/
Carbon Dioxide Ternary System . . .............. 46
6. Mole Fractions of Phenol in Water (x). Phenol in
Carbon Dioxide (y), and Equilibrium K-values . . 47
7. Calculated Values of y (Mole Fraction of Phenol
in Carbon Dioxide) at 0.33 and 0.803 g/cc . . . . 52
8. Values for Calculating the Number of Overall
Gas-phase Transfer Units ....................... 61
'9. Summary of Co-current Extraction Run Data . . . . 64
10. Converted Co-current Extraction Run Data . . . . 66
11. Summary of Phenol Loading Capability of Carbon
Dioxide at 40 Degrees Celsius . . . ............ 72
ix
LIST OF FIGURES
FIGURE PAGE
1. Reduced Pressure-Density Diagram Showing
Supercritical Fluid (SCF) and Near-Critical
Liquid (NCL) Regions . . . .......... . . . . . 5
2. Schematic of Counter-current Extraction Apparatus
as Originally' Designed and B u i l t ......... 20
3. Schematic of Workable Counter-current Extraction
Apparatus................................... . 24
4. Schematic of Concurrent Extraction Equipment . . 25
5. Phenol/Water/Carbon Dioxide Ternary Phase
Diagram........................................ 40
6. Experimental Pressure and Density Diagram for
Carbon Dioxide at 40 Degrees Celsius . . . . . . 42
7. Three Dimensional Pressure-composition Plot of
Phenol/Water/Carbon Dioxide System at 40 Degrees
Celsius, Rotated 45 Degrees . .................. 48
8. Three Dimensional Pressure-composition Plot of
Phenol/Water/Carbon Dioxide System at 40 Degrees
Celsius, Rotated 135 Degrees .................. 49
9. Phenol/Water/Carbon Dioxide Equilibria Measured
Near.1200 and 2400 psi, at 40 Degrees Celsius . . 51
10. Calculated Values of x (Mole Fraction of Phenol
in Water) and y (Mole Fraction of Phenol in Carbon
Dioxide) for the Phenol/Water/Carbon Dioxide
System, for Varying Densities, at 40 Degrees
Celsius................................ . 53
11. Graphically Determined Values of Carbon Dioxide
Phase Density and Composition for the Phenol/
Water/Carbon Dioxide System, for Varying Values
of x (Mole Fraction of Phenol in Water), at
40 Degrees Celsius 55
X12. Calculated Distribution Coefficients for Varying
Values of Carbon Dioxide Density, at 40 Degrees
Celsius..................................... 56
13. Flow Schematic for Counter-current Extraction
R u n ......................................... 57
14. Working Curves for Obtaining Gas-phase
Composition Values .............................. 58
15. Graphical Integration of the Number of Overall
Gas-Phase Transfer Units .............. . . . . 62
16. Normalized Extraction Efficiencies and Residence
Times for Co-current Phenol/Water Extraction
Experiments at 2400 psi and 40 Degrees Celsius . 68
17. Normalized Extraction Efficiencies and Residence
Times for Co-current Phenol/Water Extraction
Experiments at 2800 psi and 40 Degrees Celsius . . 69
18. Normalized Extraction Efficiencies and Residence
Times for Co-current Phenol/Water Extraction
Experiments at 1500, 1800, and 2000 psi and
40 Degrees C e l s i u s ......................... 70
19. Phenol Extraction Capabilities of Carbon Dioxide
with' Respect to Pressure, at 40 Degrees Celsius . 74
20. Summary of Normalized Extraction Efficiencies and
Residence Times for All Co-current Experiments
Using Phenol/Water Standards ................... 75
21. Equations of Best Fit for Normalized Extraction
Efficiencies and Residence Times for Co-current
Phenol/Water Extraction Experiments at 2400 and
2800 psi and 40 Degrees Celsius . ................ 77
22. Equation of Best Fit for Normalized Extraction
Efficiencies and Residence Times for All
Co-current Experiments Using Phenol/Water
Standards, at 40 Degrees Celsius . . . .
FIGURE PAGE
78
xi
23. Plot Showing Values for Normalized Extraction
Efficiencies and Residence Times for All
Co-current Experiments Using Phenol/Water
Standards, at 40 Degrees Celsius Differentiated
by Test Pressure............... 79
24. Equation of Best Fit for Normalized Extraction
Efficiencies and Residence Times for Experiments
Using Coal Gasification Wastewater, at 40 Degrees
Celsius................. 81
25. Plot Showing Values for Normalized Extraction
Efficiencies and.Residence Times for All Co-current
. Experiments, at Degrees Celsius, Differentiated by
Phenol/Water Standards and Coal Gasification
Wastewater ........................... 83
FIGURE . PAGE
xii
ABSTRACT
Existing chemical extraction techniques for the removal
of phenols from coal gasification wastewater require a
moderately expensive solvent and additional distillation
equipment for solvent recovery. This research was designed
to determine if supercritical carbon dioxide, a non-toxic,
non-flammable, and relatively inexpensive byproduct of coal
gasification, could be used for. phenol extraction.
Experiments were run to investigate phase equilibria
associated with the two phase system consisting of water,
phenol, and carbon dioxide. System pressures varied
between 1100 and 3000 psi at a temperature of 40 degrees
Celsius. Phenol concentrations in the water phase varied
from 600 to 8500 mg/1. Phase equilibrium data were first
used to determine the ability of supercritical carbon
dioxide to extract phenol from water, and later to
determine the extraction column characteristics.
Extraction equipment was built to test the feasibility
of continuous extraction of phenol from water and
wastewater. Twenty-three successful experiments were run,
including one counter-current and seventeen co-current
experiments using phenol/water standards of known initial
concentration, and five co-current experiments using
wastewater. The best extraction capability was obtained
using counter-current extraction techniques.
Extraction efficiencies, based on the approach to
theoretical equilibrium, were determined for each of the
co-current experiments. These efficiencies ranged from
81% to 145% for the phenol/water standard experiments, and
from 76% to 163% for wastewater experiments. Relative
extraction efficiencies were based on Run No. 5, which
had the highest efficiency of a phenol/water standard
experiment (145%). Extraction of phenol and phenol
derivatives from coal gasification wastewater appeared to
be higher than would be expected, based on equilibrium
data. This could be due to additional experimental error
involved in measuring the original phenol concentration in
the wastewater. Also, the existence of co-solvents in the
wastewater could enhance the solvent power of supercritical
carbon dioxide.
IINTRODUCTION AND PREVIOUS RESEARCH
Introduction
One of the major engineering challenges of this
century will be the conversion from an energy system based
on naturally occuring crude oil to a system based on the
production of synthetic fuels. Future fuels will consist
of hydrocarbons, alcohols, and hydrogen produced mainly
from coal, but also from lignite, peat, biomass, and
eventually from genetically programmed organisms utilizing
sunlight.
In the production of synthetic fuels from coal and
lignite, one of the main goals of most existing technology
is the production of carbon monoxide and hydrogen, commonly
called synthesis gas. ■ These chemical building blocks are
then combined in a variety of ways to form a wide spectrum
of hydrocarbons and alcohols. However, in the production of
the synthesis gas, there are many side reactions involving
steam, elemental carbon, and the complex mixture of mole
cules that comprise coal and lignite. Given the appropriate
reactor residence times, temperatures, and pressures, a
family of prominent side reactions are associated with any
particular type of conversion facility.
2The Lurgi coal gasification reactor, one of the older
and more proven technologies, produces a significant amount
of water soluble chemicals, including phenols, phenol
derivatives, and ammonia. Data obtained from the Grand Forks
Slagging Gasifier at the Grand Forks Energy Research Center
indicate that this type of reactor typically produces
wastewater streams containing up to 8000 parts per million
(ppm) phenol and phenol derivatives, and ammonia
concentrations approaching 14,000 ppm. As a result
considerable effort and expense are required to remove
these components from process water. These concentrations
are toxic to micro-organisms commonly used in wastewater
treatment, resulting in designs relying on chemical
treatment. Standard liquid-liquid extraction of phenols
from water with isopropyl ether requires a moderately
expensive solvent and the addition of extra distillation
equipment for solvent recovery.
There are distinct advantages for finding another
solvent that would eliminate the need for the additional
separations equipment. Carbon dioxide appears to be an
economically attractive solvent because it is a readily
available coal conversion byproduct and could be used on a
once-through basis without recovery. Fortunately, carbon
dioxide is also an effective solvent for supercritical fluid
extraction (SCFE) and has been used to remove organics from
3many solid mixtures. Applications of SCFE using carbon
dioxide include isolating valuable constituents from natural
materials like oil seeds, hops, and spices and removing or
reducing levels of undesirable substances such as nicotine
or caffeine. However, there are few, if any, commercial
applications where supercritical carbon dioxide is used to
extract components from a liquid.
Supercritical Fluids
All substances possess a critical temperature, above
which the liquid phase cannot exist. At the critical
temperature the pressure that must be applied to cause
condensation is called the critical pressure. Alter
natively, the critical pressure can be regarded as the vapor
pressure of the liquid at its critical temperature (I).
When a liquid is heated in a sealed container, the
properties of the liquid and vapor approach one another
near the critical temperature. Below the critical
temperature there are two distinct phases. As the sample
is heated through the critical temperature, the meniscus
separating the two phases disappears and one phase
results. This phase is referred to as a supercritical
fluid. Although the fluid has properties of both a liquid
and a gas it is in reality neither a gas nor,a liquid.
4At supercritical temperatures a substance that is
gaseous at ambient conditions will begin to behave like a
liquid when the pressure is raised sufficiently. Below
Sl0C, for example, liquid carbon dioxide can be used to
dissolve natural oils and a wide range of other animal and
vegetable materials. At. temperatures greater than 31°C
carbon dioxide does not liquify, but it.still can be used
as a solvent if the pressure is raised sufficiently for its
density to become similar to that of a liquid (2).
The supercritical fluid (SCF) region for a pure
component is generally defined as that region of
temperatures and pressures greater than or equal to the
critical temperature and critical pressure of the pure
component. For practical considerations, the SCF region of
interest is defined less rigorously here to include the
conditions bounded approximately by 0.9
1.0. In this region the SCF is highly compressible,
as illustrated by de Filippi (3) in Figure I. For example,
at a constant T of 1.10, increasing pressure from
Pr < 1.0 to Pr > 1.0 significantly increases the
density from relatively low values to liquid-like
densities. At higher reduced temperatures, the pressure
increase required to produce an equivalent density increase
becomes greater. This practical consideration sets the
upper bound on temperature. At a constant P of 1.50,
Re
du
ce
d
Pr
es
su
re
5
Normal
liquid
density
Reduced Density
Taken from de Philippi (3).
Figure I. Reduced Pressure-Density Diagram Showing
Supercritical Fluid (SCF) and Near Critical
Liquid (NCL) Regions.
6decreasing, temperature has a similar effect on density, and
at higher reduced pressures, the.density is less sensitive
to temperature changes. In the vicinity of the critical
point,, large density changes can be produced with either
relatively small pressure or temperature changes (4).
The use of supercritical fluids as solvents for
extraction has unique advantages. Supercritical fluids
combine the solvent power of a liquid with the transport
properties of a gas. They can be used to selectively
dissolve relatively involatile solutes.from a mixture
while allowing the simple separation of solvent and solute
by either a reduction in pressure or an increase in
temperature. Pressure is the principal processing-
variable used to alter the solvent power. In contrast,
liquid extraction solvent power is varied by adjusting the
temperature or the solvent composition; allowable
temperatures are limited by the thermal characteristics of
the solute while variations in solvent composition
increase separation and purification costs. Significant
reductions in energy requirements are also characteristics
of SCFE separations (5).
7Extraction with Supercritical Fluids
Supercritical fluid extraction (SCFE) is a technique ■
that exploits.the solvent power of supercritical fluids at
temperatures and pressures near the critical point. By
selectively choosing the appropriate solvent a wide range
of operating temperatures and pressures can be utilized to
tailor an extraction process design to a particular need.
It is possible to dissolve both solids and liquids
(including ones of low volatility) in compressed gas. The
behavior of gas/solid systems is normally simpler than that
of gas/liquid systems as the composition of the solid phase
normally remains constant. When a liquid is being extrac
ted, the situation is complicated by the influence of
dissolved gas on the properties of the liquid phase (2).
The advantages of using supercritical fluids under
pressure include the following:
High boiling components can be extracted at
relatively low temperatures. Also, at the
lower temperatures involved, heat sensitive
compounds are undamaged during the extraction
process.
Separation of the solute and solvent is
relatively easy at supercritical conditions,
usually by relatively slight pressure changes.
Separations can be achieved which are not possible
by other techniques such as distillation or
extraction.
8.
If food components are to be separated, non-toxic
solvents, such as carbon dioxide, may be used which
leave no harmful residues.
In general terms, compressed gases are relatively
cheap solvents.
The difficulties which have to be overcome include
the following:
High plant costs involving relatively high risk
capital.
Cost of maintenance normally required for high
pressure plants.
Current lack of process design data and lack of
information on equipment reliability.
In many unique and difficult processes there is
a tendency to use patent laws as a protection
against competition. The mass of patents in
the field makes it extremely difficult to
determine if infringement is likely in a
particular application.
Supercritical fluid extraction is particularly
effective in the isolation of substances of medium
molecular weight and relatively low polarity. Its
principal advantage over distillation is that separation
can be accomplished at moderate temperatures, and thus can
be applied to the recovery of heat-labile substances of
low volatility. It is therefore of interest in the food
and petroleum industries.
)
9In comparison with liquid solvents, the supercritical
fluid has high diffusivity and low density and viscosity,
allowing rapid extraction and phase separation. The
solvent power of the supercritical fluid can be varied
over a wide range by varying pressure or temperature, and
the fluid is easily recovered from the extract and from
the extraction residue as a result of its high volatility.
Again, the principal disadvantage of SCFE is the
relatively high pressures required to achieve the
necessary densities, generally in the range of 50 to 200
atmospheres. Another disadvantage is the complexity of the
phase relationships under supercritical conditions.
Theoretical techniques for predicting extraction behavior
are new and relatively untested. An extensive bibliography
assembled by McKugh and Krukonis (6) shows that most of the
modeling work in this area has been developed in the last
ten years (7-19).
The SCFE concept was first recognized by Hannay and
Hogarth (20) over a hundred years ago. They observed that
potassium iodide was dissolved in supercritical ethanol and
was precipitated on reducing the pressure. Later it was
realized that the solvent power of supercritical fluids
could be involved in geological processes through the
influence of water on rock formation, and of methane in
petroleum formation and migration. In power stations the
10
adoption of supercritical steam pressures led to the
deposition of silica on the blades of steam turbines (21).
The possibilities for substantial energy economies,
relative to normal distillation processes, have made SCFE
an area of intense interest. Arthur D. Little CS) probably
triggered this interest with the claim that supercritical
fluid extraction would make it possible to produce a gallon
of ethanol using carbon dioxide as extractant with ah
expenditure of 10,000 Btu of energy instead of the 25,000
to 50,000 Btu needed for conventional distillation.
Carbon Dioxide as a Solvent
Carbon dioxide has many characteristics which make it
suitable for use as a solvent. It is completely miscible
with low molecular weight hydrocarbons and oxygenated
organics and is therefore a good solvent for many organics
(3,22 ). It is relatively volatile, compared with most
organics likely to be ,extracted, allowing easy separation.
In addition, carbon dioxide has low viscosity and high
diffusivity, and it is totally non-toxic, non-flammable,
and relatively inexpensive (23).
. The solubility of supercritical carbon dioxide in water
is relatively small, generally less than seven-'weight
percent, even at pressures exceeding 400 atmospheres. At
near critical conditions, for example at 40°C and 80
11
atmospheres, the solubity is approximately 5.5 weight
percent (24). This allows the use of carbon dioxide for a
variety of extraction chores involving aqueous solutions.
Furthermore, carbon dioxide can be removed easily by simple
deaeration techniques.
As a supercritical fluid carbon dioxide has additional
properties that make it an excellent solvent. Its critical
temperature (31°C) and pressure (72.8 atm) are readily
accessible with well-established process technology, and
equipment. Finally, its heat of vaporization is low,
especially near the critical point, leading to low energy
requirements in many processes (3).
Table I shows a comparison of the fluid properties of
carbon dioxide as a liquid, a supercritical fluid, and as a
gas (3, 24). The ranges shown for the supercritical fluid
are those in the near critical region. At sufficiently high
pressures (approximately 500 atm at 40°C) the density of
supercritical carbon dioxide can actually exceed I g/cc.
Also, in the near critical region significant energy savings
can result due to the very low heats of vaporization.
Table 2 shows selected data relating temperature, pressure
and heat of vaporization (3). The zero value listed at the
critical point illustrates the unique quality that any
fluid possesses when the distinction between a liquid and
a gaseous phase disappears.
12
Table I. Comparison of Carbon Dioxide Fluid Properties.
Liquid Supercritical Gas
Density (g/cc) 0.5 - 1.2 0.2 - 0.7 0.001 •
Viscosity (cP) 0.5 — 1.0 0.05 - 0.10 0.01
Diffusivity (cm^/s) 10™5 IO"4 - IO"3 10"1
Table 2. Heat of Vaporization of Carbon Dioxide.
Saturation Heat of
Temperature Pressure Vaporization
F C (atm) (Btu/lb).
0 -17.8 20.8 120.1
20 4.4 38.6 95.0
60 15.6 50.9 76.6
70 21.1 58.1 63.8
80 26.7 65.9 44.8
87.8 31.0 72.7 0
Supercritical fluid extraction with carbon dioxide
has been used extensively in the food industry. Here the
properties of being non-toxic and non-flammable have
significant advantages when extracting such valuable
products as perfume oils, caffeine, hops, and vegetable
13
oils. It also is possible to remove.PCB1s from silicon-
based transformer oils and to remove adsorbed materials
from common adsorbents such as activated carbon and
synthetic resins.. Table 3 shows a number of examples of the
use of carbon dioxide for supercritical fluid extraction.
Table 3. References with Examples
Extraction Using Carbon
of Supercritical
Dioxide.
Fluid
Substance i-3 O O P, atm ref.
Vegetable oil 20, 55 150-450 3
Polychlorinated
biphenyls (PCB) 100 120-265 ■ 3
Silicone oil 100 160-250 3
Alachlor (pesticide) 50 - 120 88-272 3
Caffeine 90 160-220 21
Hops A O unidentified 25
Lilac oil 34 90 26
Lemon peel 40 300 27
Black pepper 60 100-400 27
Almond oil 40 600 27
14
Previous Research
Research done previously at Montana State University
has investigated the use of supercritical fluids for
separating organics from the undesirable mineral portion
of peat. Granlund (28) and Smith (29) investigated the
semi-continuous flow extraction process where a variety of
supercritical and near supercritical fluids were passed
through a container containing a measured amount of peat..
Organics were extracted from the mineral portion of the peat
and yields were measured for each type of solvent used.
Earlier-work performed by Myklebust (30) investigated the
potential for utilizing supercritical fluids for batch
extraction of organics from peat.
The University of Delaware Chemical Engineering
Department has been active in SCFE research. Van Leer and
Paulaitis investigated the solubilities of phenol and
chlorinated phenols in supercritical carbon dioxide (31),
and McHugh and Paulaitis investigated the solubilities of
solid naphthalene and biphenyl in supercritical carbon
dioxide (32).
Kurnik, Holla, and Reid (33) of the Massachusetts
Institute of Technology studied the solubility of several
solids in supercritical carbon dioxide and ethylene. They
15
include a table of references with the solubilities of
various solids in supercritical fluids. This data is shown
in Table 4.
Table 4. References with Solubility of Solids and Liquids
in Supercritical Carbon Dioxide.
Substance T, C P, atm ref.
Naphthalene 35, 45, 55 100-300 23
Naphthalene 35-55 60-330 34
Ethanol/Water 40, 60 75-204 3 5
Isopropanol/Water 40 102 35
Phenanthrene 40 135-544 36
Diphenylamine 32-37 50-225 37 :
Phenol 36, 60 78-239 31
Chlorinated Phenols 36 80-130 31
Napthalene 35-65 86-288 32
Biphenyl 36-58 105-484 32
Biphenyl 55 503-531 38
Naphthalene 58-80 1-250 38
16
Probably the most extensive research involving the
solubility of various substances in carbon dioxide was
published by Francis in 1954 (39). Mutual solubilities of
liquid carbon dioxide with each of 261 other substances were
reported. Nearly.half of these substances are miscible with
carbon dioxide. Francis noted some relations to structure
and plotted phase diagrams for 464 ternary systems involving
carbon dioxide. He also pointed out that carbon dioxide has
a strong homogenizing action upon pairs of other liquids at
moderate concentrations, but a precipitating action at
higher concentrations. In contrast to most solvents it has
a selectivity against dicyclic hydrocarbons. Co-solvents
were found necessary to make these unusual properties
effective in the solvent extraction of hydrocarbon mixtures.
While this paper does not include any research involving a
supercritical phase, the measurement conditions ("liquid
carbon dioxide at or near room temperature") are in the
near-critical region, and the data provides an invaluable
tool for roughly estimating the probable success one might
have using supercritical carbon dioxide with any of the same
combinations of substances. Using some of the groupings
that Francis presents might also help predict the
suitability of using supercritical carbon dioxide as an
extractive agent for similar mixtures.
17
RESEARCH OBJECTIVE
The purpose of this research is to investigate the
feasibility of using supercritical carbon dioxide to
extract phenols from coal gasification wastewater. In order
to achieve this ultimate objective the work presented in
this paper focuses on the development and testing of the
necessary extraction equipment, investigation of water/
phenol/carbon dioxide equilibrium data, and the development
of procedures for analyzing phenols in dilute aqueous
solutions.
. Extraction equipment was ordered, assembled, tested,
and modified to obtain phenol extraction data at approxi
mately 40°C and pressures ranging from 1100 to 2800
psig. Efforts were made to obtain data representing
continuous flow conditions, both counter-current and
co-current. These data are compared with the equilibrium .
data, obtained as part of this research, to evaluate the
effectiveness of each set of extraction conditions.
18
MATERIALS, EQUIPMENT, AND PROCEDURE
Materials
Carbon dioxide was obtained as a technical grade
liquid, heated, and compressed into a special bottle rated
at 6000 psi. A Haskel Model.AGD-62-C Gas Booster
compressor was used to boost the CO^ from approximately
850 psi to 5000 psi.
Original extraction data was obtained using a 10,000
mg/1 reagent grade phenol in distilled water standard
solution. After the extraction equipment had been
sufficiently tested and operating parameters better
understood, another type of solution was tested for
extraction. This solution was obtained from the Grand
Forks Energy Research Facility and consisted of process
water condensed from their experimental slagging gasifier.
Equipment
Equilibrium Data Equipment
Equipment used for generating phase equilibrium data
included a Parr Instrument Company Series 4000.pressure re
action apparatus. This 500 milliliter bomb was constructed
19
of Inconel Ibl-S alloy. The bomb was specially fitted
with a modified head to allow two valves for filling with
the appropriate phenol/water solution and pressurizing with
carbon dioxide. One of the valves was.also used for
sampling the liquid phase. The apparatus also included a
rocker equipped with an electrical heater which was used to
keep the fluids thoroughly mixed at 40°C. The bomb
temperature was monitored using a thermocouple located at
the base of the bomb and connected to a 12 channel temper
ature indicator. Temperature was manually controlled by a
Fisher Scientific rheostat which supplied power to the
rocker heating coils.
Extraction Apparatus
The equipment used in this research was designed to
facilitate extracting phenol from water on a continuous
basis. Figure 2 shows a schematic of the original extrac
tion equipment. The carbon dioxide and water feed systems
were essentially unchanged for the duration of all the
extraction runs. All of the significant equipment changes
involved the column itself and solution and solvent
discharge lines. Essentially all the high pressure lines
used in this research were located behind a three-sided.
Wet test meter
Discharge carbon
dioxide
Surge
check
valve Phenol
accumulator
Phenolic Measuring
wastewater burrette
Safety f]--
Carbon
dioxide
-.fillJ Porc stripped
water
vent
Check
valve
Nitrogen
fill port
Nitrogen vent
Extraction column,
Heated enclosure
Heated, high pressure
carbon dioxide feed tank
Water pump
High pressure
BPR - Back pressure regulator
TI - Temperature indicator
PI - Pressure indicator
nitrogen storage
FI - Flow
MO
Figure 2. Schematic of Counter-current Extraction Apparatus as
Originally Designed and Built.
21
high-strength.steel shield. Control valves, gages, and
electronic meters were accessible from the front of the
shield.
Carbon dioxide was stored under sufficient pressure to
provide the appropriate test pressure for each run. The
Haskel Gas Booster compressor was used to transfer. CC^
from a standard cylinder to a special high pressure
cylinder rated at 6000 psi. The high pressure cylinder was
warmed to supercritical temperatures (approximately
40°C) using an electrical heating blanket. The
temperature was manually controlled with a Powerstat Type
3PN1168 variable transformer, and monitored using a
thermocouple connected to a Cole-Palmer Model 8530 12
channel electronic indicator. Carbon dioxide pressure was
regulated by a Tescom Model 44-1126-24-013 back pressure
regulator. Heat was supplied with a 100 watt electrical
heat tape as the pressure was reduced to provide heat of
vaporization and to further raise the temperature to
approximately 40°C. The temperature was measured with
a thermocouple attached to the regulator surface and
monitored using the Cole-Palmer indicator. From the
regulator the solvent then entered a system of control and
check valves which provided flow control to the extraction
column. Once the carbon dioxide had passed through the
column its pressure was reduced to atmospheric using a
22
Grove Mity-Mite Model S-91XW back pressure regulator. Back
pressure was supplied by a Parr Instrument Company Series
4000 pressure reaction apparatus filled with nitrogen up to
5000 psi. This 500 milliliter bomb was constructed of
Inconel 757-S alloy. Once again the.solvent stream was
heated with electrical heat tape (200 watts) while its
pressure was reduced. This was to prevent icing (both
water and CO2) and to help prevent the deposition of
solid phenol in the regulator and exiting tubing. Both
heat tapes were manually controlled with a Powerstat Type
3PN1168 variable transformer. Atmospheric carbon dioxide
was then passed through a Parkinson Cowan wet test meter to
measure the actual solvent flow rate. The CO2 ports
were designated as inlet/outlet ports since they were used
as such depending upon whether the particular extraction
was counter-current or co-current.
The phenol/water solution was stored in a stainless
steel Griffin beaker. Solution was gravity fed from the
beaker into a Milton Roy Model HDB-1-30R pump capable of
delivering a maximum of two milliliters per minute at 3000
psi. The flow rate was measured by timing the removal of
solution from a burette and valve apparatus as shown in
Figure 2. The solution was then pumped through 1/4 inch
stainless steel tubing past a check valve, a rupture
disc, two pressure indicators, and into the extraction
23
column. The pressure was measured by a Model DPS 0081
electronic pressure sensor made by Autoclave Engineers and
by a 5000 psi gage used as a cross-check. After pressure '
measurement the solution lines entered an oven containing
the extraction column. The 1/4 inch lines were then
reduced to 1/8 inch before actually entering" the column
through a side port located near the top of the cylindrical
portion of the column. The water entry port was located
approximately one inch below the COg inlet/outlet port.
Water and COg exiting the column were originally
reduced to atmospheric pressure using the same type of back
pressure regulator. Both back pressure regulators were
connected to the same nitrogen pressure system, as shown in
Figure 2, so that the entire column and the fluids flowing
through it would sense the same conditions. However, the
slight differences in the regulators were enough to
prohibit the use of this arrangement. The column length
between the water inlet and the water outlet, determined by
the regulator position, was limited.to approximately 20
inches, or a pressure head of less than one psi. At system
pressures between 1100 and 3000 psi the back pressure
regulators just could not provide the necessary control.
The extraction equipment was then modified, as shown in
Figure 3, so that water exiting the column would enter the
bottom of a bomb that was pressurized by a "vapor" line
24
Pressure
equalization
vessel
<■
Carbon dioxide
at atmospheric
— > pressure
Water/phenol mixture
Carbon
dioxide Vv
Extraction column
(in heated enclosure)
Pressure lock
sampling system
BPR - Back pressure regulator
Figure 3. Schematic of Workable Counter-current
Extraction Apparatus.
from the top of the column. Extracted water accumulated in
the bomb until the run was considered complete. A pressure
lock system of valves allowed the removal of approximately
four milliliters of extracted water for sampling purposes.
A final major equipment modification was made to enable
co-current extraction runs. A schematic of this arrangement
is shown in Figure 4. Here, the pressure equalization
vessel doubles as a phase separator and vessel to store
25
Discharge
^ carbon
dioxide
Carbon
dioxide Water/phenol mixture
Phase Extraction vessel
separation (in heated enclosure)
vessel
Pressure lock
sampling system
BPR - Back pressure regulator
Figure 4. Schematic of Co-current Extraction Equipment.
extracted solution until the end of the run. Valves at the
bottoms of the settler vessel and extraction column were
added to enable each system to be isolated while lines were
being purged and sampled.
The extraction column itself was totally enclosed in a
steel box which was fitted with two 500 watt electrical
heaters manually controlled by a Powerstat Type 3PN1168
26
variable transformer. Oven temperature was measured by a
thermocouple and monitored by the Cole-Palmer indicator.
The column was 12 inches long with an inside diameter of
0.688 inches and outside diameter of one inch. The top end
of the column was fitted with a custom-machined steel
fitting having three ports. The top fitting had three
ports, a carbon dioxide inlet/outlet port, a water solution
inlet port, and an unused port. The bottom of the column
was fitted with an Autoclave Engineers tee having two
ports, one for the carbon dioxide inlet/outlet port, and
the other for the water outlet. All four inlet and outlet
ports were reduced to 1/4 inch NPT openings. Three of
these openings were reduced further to 1/8 inch Swagelok
fittings to facilitate tubing connections in the close
quarters available in the oven. The fourth opening,
located at the lowest point, was used for the extracted
water solution outlet. This port provided a straight line
path for the exiting water solution and was left as large
as possible to reduce any flow restrictions.
Three types of column packing were used in separate
trials. These included 1/4 inch glass Raschig rings, 1/16
inch glass helices, and 1/16 inch steel helices., Support
screens were cut to fit the column ends to prevent packing
from entering any of the inlet/outlet ports.
27
Analytical Equipment
Phenol concentrations were measured using a Varian
1800 Series Gas Chromatograph with a flame ionization
detector. A new custom packed glass column, specifically
designed for phenol and phenol derivatives analyses, was
obtained and preconditioned according to the manufacturer's
specifications. The column packing consisted of KG-02 on
Uniport HP, with 80/100 mesh. Column length was ten feet
with a two millimeter inner diameter and 0.125 inch outer
diameter. All the phenol analyses, for this project used
this single column at an oven temperature of 175°C,
with a helium carrier gas and a hydrogen flame.
While this research did not involve the removal of
ammonia or chemical oxygen demand, it was anticipated that
future work would. Equipment was therefore organized for
the analysis of these two parameters. An Orion 901 Ion-
alyzer with an ammonia ion-specific electrode was prepared
to measure ammonia concentrations. The complete procedure
and specific reagents used in the ammonia analyses are
presented in Appendix A. Chemical oxygen demand standards
were measured using a large hot plate, a thermometer
reading at least 200°C, 500 milliter Erlenmeyer flasks,
and reagent grade chemicals including mercuric sulfate,
ferroin indicator, ferrous ammonium sulfate, concentrated
sulfuric acid, and rapid dichromate acid solution. The
special reagents and procedures used in this method are
described in Appendix A.
Experimental Procedures
Phase Equilibrium Run Procedures
A known volume, usually 100 milliters, of a phenol/
water standard was poured into the Parr bomb. The standard
consisted of a known concentration of phenol, either 625,
1250, 2500, 5000, or 10,000 mg/1. Carbon dioxide was then
compressed into the bomb to a pressure of about 1000 psi
for the low pressure tests and to about 1500 psi for the
high pressure tests. The bomb was then placed in the .
rocker (but not rocked) and heated to 40°C to obtain a
pressure reading. Carbon dioxide was added or vented as
needed to attain the desired system pressure. The rocker
was then set in motion and the heat adjusted so that the
system would stay at 40°C for at least 10 minutes.
After the system had stabilized the rocker was stopped and
the bomb placed in an inverted position while the
temperature was maintained. ' The liquid phase and the
supercritical fluid phase were allowed to separate for
about five minutes and a small amount of fluid was allowed
to be vented. Immediately after the initial venting a
three milliliter liquid sample was taken for analysis.
28
29
Extraction Run Preparations
In preparation for each run the high pressure carbon
dioxide cylinder was checked to see if there was sufficient
solvent pressure for the particular run and if the bottle
temperature was approximately 40°C. Powerstats were set
to approximate settings to maintain the carbon dioxide
temperature across both pressure drops and to maintain the
oven temperature at 40°C. Experience was the best guide
as to what settings would be required. If the high
pressure cylinder needed to be recharged the Haskel
compressor intake was connected to a liquid CO2 cylinder
at room temperature, the discharge was connected to a port
located on the front of the panel, the appropriate valves
were opened and closed, and the compressor was started.
When the high pressure cylinder was charged to about 4500
psi the valves, were reset and the compressor was
disconnected.
Also, before a run was begun, the back pressure
regulator system was set to the desired system pressure and
a check made for an essentially perfect seal. A nitrogen
leak during a run of any reasonable length would cause the
system pressure to change and ruin an otherwise good run.
Once the original equipment was in place the nitrogen
system was very reliable. Occasionally the high pressure
30
nitrogen bomb would also need to be recharged due to manual
adjustments in the pressure settings. The Haskel
compressor intake was then connected to a nitrogen cylinder
and the discharge to a special nitrogen port on the front
of the panel. Each back pressure regulator was isolated
from the bomb, and the bomb was then charged with up to
5000 psi of nitrogen. The compressor was then disconnected
and the back pressure regulator system was charged with
whatever pressure was.desired.
The phenol/water solution was checked before and
during each run to ensure that the pump would not run dry
and .to provide a continuous feed to the column. All but
the last five runs were charged using a 10,000 mg/1 phenol
standard prepared in- the lab. The last five runs were
charged with actual coal gasification wastewater containing
approximately 5000 mg/1 phenol, a variety of cresols,
ammonia, and other organics.
The wet test meter was checked before each run for
the proper water level and proper gas seal. The carbon
dioxide discharge line was vented into a 500 milliliter
flask to allow any solids or liquids to drop out before
entering the wet test meter. The flask was checked for
a proper seal.
31
Extraction Run Procedures
Although there were several different types of
equipment arrangements, most of the experimental procedures
were essentially the same. The principal differences
involved the collection of extracted solution samples. At
the start of each run the Tescom back pressure regulator
was adjusted to approximately 50 to 100 psi higher than the
system pressure. ■ The carbon dioxide feed control valves
were then used to fine-tune the solvent flow rate for each
particular run. This flow rate was one of the principle
changeable parameters for each run and, as it turned out,
was one of the biggest sources of problems, in producing
reliable extraction data. Once solvent flow was reasonably
established the water solution pump was started and the
flow rate checked and adjusted if necessary. Usually the ■
solvent flow would then need to be readjusted to fine-tune
the balance between solution and solvent flow.
After both solvent and solution flows were established
at reasonally constant rates the time and the wet test
meter reading were recorded. Periodic flow checks were
made to ensure that the flows remained constant.
Adjustments were made as necessary to keep the system as
close to a steady-state as possible. Almost always the
only adjustment necessary was the adjustment to the carbon
32
dioxide flow. This was due to the constantly varying
pressure on the high pressure side of the Tescom back
pressure regulator, which changes the cooling effect across
the regulator and the subsequent heat requirement of the
manually controlled heat tape. Slight changes in original
carbon dioxide pressure required periodic control valve
adjustments as the run progressed.
Once the run was considered to have reached steady-
state (approximately two to six hours, depending on carbon
dioxide mass flow rate) the solvent and solution flows were
stopped and the extracted solution was sampled. This
allowed a sufficient amount of solution to accumulate in
the bottom of the collection bomb to enable a representa
tive sample to be taken. Valves located at the bottom of
the column and at the bottom of the collection bomb were
closed immediately after the flows were stopped. The system
was then allowed to rest, undisturbed, for approximately
five minutes to allow any phenol-laden (pregnant) carbon
dioxide that was undissolved in the water phase to float to
the surface of the water phase. The sample line between the
two valves was then drained via the lock valves. The valve
below the collection bomb was then reopened to allow the
liquid in the bottom of the bomb to flow into the sampling
line. Liquid samples were then removed, approximately 3.5
milliliters at a time, and collected in labeled sample vials
33
Originally, each vial was analyzed to get some idea
of whether the runs were indeed long enough and to check
if the concentrations of phenol changed significantly with
each sample. It appeared that the first flush, which
included both solution and solvent, was not representative.
Actually, that result was expected for co-current flow
conditions since any phenol dissolved by the carbon dioxide
would be redeposited when the combined mixture was reduced
to atmospheric conditions. Also, the first sample after
reopening the bomb bottom valve was questionable since
carbon dioxide may have been trapped in some of the tubing
immediately below the bomb. This sample analysis was
normally discarded. Once several runs had been analyzed it
appeared that the first line flush should be discarded
along with at least the first three to four milliters of
solution in the bomb. Then, approximately 10 milliters of
extracted solution would be collected in a graduated
cylinder and about three to four milliters transferred to a
labeled vial for later analysis. These samples were
considered representative of the steady state conditions
during each run.
The last bit of liquid to be sampled from the bomb was
generally accompanied by a considerable amount of carbon
dioxide and was considered to be non-representative of
(
34
steady-state conditions. This sample, was discarded, and
generally the analysis of the one previous sample was
considered questionable if it varied significantly from the
average of the earlier samples.
Sample Analyses
Phenol concentrations were measured using a Varian
1800 Series Gas Chromatograph (GC) with a flame ionization
detector. All the phenol analyses for this project used
the packed glass column, specifically designed for phenol
and phenol derivatives in a polar solvent (water). The
helium flow rates were varied as needed to maintain
reasonable peak sharpness, peak separation,and analysis
time. Normal helium flow rates corresponded to regulated
helium pressures of 25 to 40 psig supplied to the column.
Air and hydrogen were adjusted to approximately a 10:1
flow ratio sufficient to maintain a stable flame. The.
attenuation and range were adjusted to attain a peak height
suitable for the particular phenol concentration range in
question, ideally a peak height approximately 80% of full
scale. A prepared phenol standard was then injected to
determine a basis for unknown sample measurement.
Typically, a sample size of approximately one microliter
was used and enough repetitions were made of standards and
samples to account for slight variations in sample sizes.,
35
A concerted effort was made to use the same sample size for
both standard and unknowns to eliminate any unnecessary
measurement errors due to detector sensitivity, column
overloading, or other parameters.
Since even the highest phenol concentrations were only
one percent (10,000 mg/1) phenol, very sensitive settings
were required. Even so, the peaks were not ideal, often
having wide bands with trailing tails. In order to be
consistent the areas under the top half of each peak were
measured. Unknown sample concentrations were calculated
based on a directly proportional comparison of the average
area of the standards. For example, an unknown sample peak
with an area of 1.2 times the area of the standard's peak
would have a concentration 1.2 times that of the standard, .
assuming the sample sizes were the same and the
attenuation, range, and all other parameters were constant.
Helium flow rates were kept constant, as much as practical,
during any particular analytical run, and only changed to
adapt to the measurement of new phenol standards.
A few problems were encountered due to sample sizes.
Excessive sample sizes, generally greater than two
microliters, sometimes caused wide peaks with excessively
large tails, off-scale problems, and carrier gas pressure
(flow) swings. Too small sample sizes created problems
with getting consistent sample sizes, often resulting in
36
peaks with widely varying heights and areas. Also, if
the unknown sample's concentration was significantly lower
than the standard's concentration, a short, wide peak was
produced and a reliable comparison was not possible. The
operating parameters of the GC,. the sample sizes, and the
phenol standard would then be.varied until a suitable
combination was found, and analysis would be continued.
All samples, both extraction run and equilibrium
phase, were allowed to cool to approximately room
temperature principally to allow carbon dioxide bubbles to
separate from the liquid but also to allow the liquid
density of each analyzed sub-sample to be virtually
identical. Sub-samples were drawn from the sample vial
with a 10 microliter syringe and injected into the GC
injection port. Every effort was made to keep the detector
clean and to keep the hydrogen, helium, and air flows as
constant as possible.' Extreme care was taken when
analyzing the equilibrium phase samples. Approximately,
five GC analyses of each sample were averaged and two
standards were run before and after each sample group.
Standards were chosen so that phenol concentrations in both
the standards and the samples would be nearly the same and
would register between 50% and 95% of full scale on.the
recorder chart. . Phenol concentations in the carbon dioxide-
rich phase were calculated based on a mass balance of the
37
amount of phenol left in the water phase (measured), the
original amount of phenol in the water, and a calculated
amount Of carbon dioxide placed into the bomb. The actual
concentration of phenol in the supercritical phase was not
directly measured. Normally, duplicates of extraction run
samples were analyzed and an average of the two was
considered to be sufficiently accurate. For some of the
later runs, when it was clear that several samples from the
same run were all representative of the extraction run,
individual analyses were grouped together and averaged for
a nominal value. This helped reduce the excessive amount
of time required for data acquisition and interpretation.
38
EXPERIMENTAL RESULTS AND DATA ANALYSIS
Introduction
Altogether, fifty-three extraction runs were attempted
including forty-eight runs using 10,000 mg/1 phenol/water
standard and five runs using actual coal gasification
wastewater. Of these runs, twenty-three were successfully
completed, one counter-current and the remainder co-current
In addition there were seventeen successful phase equili
brium runs at pressures ranging from 1100 to. 2 80 0. psi and
initial concentrations ranging from 625 to 10,000 mg/1
phenol.
/
The initial intent of this research was to test the
feasibility of using counter-current extraction techniques
with the two phase, three component system of phenol, water
and supercritical carbon dioxide. The extraction equipment
was designed and built with the expectation that a number
of different parameters could be tested to determine some
optimum operating conditions. These parameters included
pressure, relative flow rates of solvent and liquid
solution, temperature, and column packing. Unfortunately,
only one counter-current extraction run was successfully
completed after twenty unsuccessful tries. Nine more
39
counter-current flow attempts failed, before it was decided
to try co-current extraction tests. The most frequent
cause of run failure was the inability to Control the
carbon dioxide flow within the required degree of.accuracy.
Reliability of Data
Throughout this research there are two general
categories, one relating to extraction runs and one
relating to phase equilibrium data. A discussion of data
reliability must necessarily separate the two, since there
is a distinct difference in reliability.
Originally, the phase equilibrium data was gathered
to get some idea of the feasibility of extracting phenol
from water using supercritical carbon dioxide. Very
little information was available in the literature to
indicate what would happen with this system or at what
conditions an extraction might be optimal. The liquid-
liquid-solid ternary phase diagram published by Francis
(39), and enlarged in Figure 5, shows that the region of
interest, for this research, lies entirely within one
percent phenol of the water-carbon dioxide baseline. An
attempt was made to experimentally develop a set of
equilibrium data for this ternary system at the dilute
conditions in question.
40
Carbon Dioxide
One liquid phase
Diagram taken from Francis (39),
representing data at approximately
26 degrees Celsius and 65.1 atm.
Solid, \
\ Liquid,
\ Liquid
Liquid, Liquid, Liquid
Liquid,'
\ Liquid
One
liquid
phaseLiquid, LiquidOne
liquid
phase
PhenolWater
Figure 5. Phenol/Water/Carbon Dioxide Ternary Phase
Diagram.
41
Since the. solubility of phenol in supercritical carbon
dioxide is a strong function of density and the facilities
to analyze high pressure carbon dioxide directly were not
available, it was necessary to use the mass balance approach
in measuring the phenol concentration in the supercritical
fluid (SCF) phase. The actual mass of carbon dioxide can be
determined by knowing the volume available to the SCF. phase
and by knowing the density of carbon dioxide at 40°C and
the test pressure. Density data used for these
calculations were taken from the temperature-entropy
diagram in Perry and Chilton (24). A plot of density
versus pressure is shown in Figure 6.
Water samples taken from each equilibrium data
experiment were carefully analyzed by gas chromatograph to
determine the concentration of phenol in the water rich
phase. Using the published experimental density data and
the GC phenol concentrations, a phenol concentration was
calculated for the SCF phase and expressed as moles phenol
per moles carbon dioxide. A number of possibilities
existed for the introduction of error in this procedure.
Major concerns included the following:
I. GC concentrations were considered to be accurate
to within 10% of the actual value. Any error in
analysis was compounded by the use of mass
balance for determining the mass of phenol in
the carbon dioxide phase.
Ca
r
bo
n
Di
ox
id
e
D
en
s
it
y
(g
/c
c)
42
Data taken from Perry and Chilton (24)
300020001500
Pressure (psig)
Figure 6. Experimental Pressure and Density Diagram for
Carbon Dioxide at 40 Degrees Celsius.
43
2. The carbon dioxide density, and therefore its
mass, was determined by interpolation of the
data available from the literature.
3. Since the data was gathered near the critical
temperature of carbon dioxide, slight changes
in temperature would result in relatively large
changes in density. The large amount of mass
involved in the test equipment provided a
considerable amount of lag time in heat input
and temperature change and the temperature
measuring apparatus provided a digital readout
accurate only to the nearest degree Celsius.
Therefore it is difficult to determine the
actual margin of error due to measurements that
were slightly higher or lower than 40 degrees
Celsius.
4. Although a rocker was used to mix the phenol
■solution and the carbon dioxide, there was
no definite way of knowing whether an
equilibrium had been reached. As discussed
below, residence time was a critical parameter
in the extraction runs and could possibly
have been a significant factor in the
accuracy of the equilibrium data.
The principal concern with the phase equilibrium data
is the dependence of an uncertain mass balance to calculate
the phenol concentration in the SCF phase. It would be far
better to be able to actually measure the concentration in
both phases.. The equilibrium data does, however, provide a
general idea of what to expect during co-current extraction
runs, where the approach to equilibrium should be the upper
limit. For counter-current extraction the availability of
equilibrium data allows the calculation of the number of
theoretical trays in a column and the height of an equiva
lent theoretical plate for a particular set of parameters.
44
Equilibrium data obtained from this research was used to
roughly analyze both the co-current and the counter-
current extractions.
Several kinds of data were obtained from the extrac
tion experiments. These included solution and solvent flow
ratesf temperatures, pressures, and phenol concentrations
determined by GC analysis. Solution flow rates were
measured very accurately, to within 0.5% of the actual
flow, and were subject to only slight variation during any
of the valid runs. Solvent flows, on the other hand, were
much harder to control, and varied during some runs by as
much as 20% above and below the average flow. Since the
carbon dioxide flow rates were measured by a wet test
meter, which accumulated the total volume of CC^ during
a timed run, an average flow rate was used for analyzing
the extraction data. Oven (and column) temperatures were
manually controlled to within one degree Celsius of the
desired 4O0C. System pressure was generally kept
within 25 psi of the run's nominal pressure. Extraction
runs involving very high rates of carbon dioxide flow
usually generated problems with pressure drifting due to
excessive cooling near the back pressure regulator. On the
last run this problem was virtually.eliminated by
rearranging the electrical heat tape to more directly heat
45
the back pressure regulator itself. Much higher flow rates
were then made possible without causing icing problems.
Phenol/water standards were prepared in the lab and
were accurate to within 0.02% of the nominal 10,000 mg/1.
This is much more accurate than the estimated accuracy of
any one GC analysis, considered to be within 10% of the
actual concentration. However, known standards were
analyzed before and after any unknowns, and unknown samples'
were usually analyzed at least twice to minimize any
analytical error. By using the GC peak area averages of
the standards for reference, and the peak area averages of
unknowns for the final concentration, it is believed that
the actual concentrations measured and reported in this
paper are accurate to within 10% of the true values. The
final five extraction runs, using actual coal gasification
wastewater, were analyzed on'a relative basis. Since there
were other phenol derivatives present in this solution,
there was some interferences present in the GC charts
involving a widening of the peak and secondary peaks riding
the main phenol peak's tail. Unknowns, that is, the
extracted solution, were analyzed at the same GC setting as
the original solution, and the results were expressed as a
percentage of the original. Due to the wide differences in
peak heights and widths, the accuracy of this data is
probably on the order of plus or minus 20% of the actual
46
minus 20% of the actual measurement. In all extraction
runs, if any data were considered to be out of the above
ranges, the run was considered unusable and the data were
not used.
Phase Equilibria
Raw phase equilibrium data are shown in Table 5.
These data were then converted to convenient units for
comparing concentrations, that is, moles phenol per mole
of solvent.
Table 5. Raw Phase Equilibrium Data
Phenol/Water/Carbon Dioxide
for the
Ternary System.
Pressure Phenol Concentrations.(mg/1)
Run (psig) Initial Final
I 1,210 625 597
2 1,205 1,250 1,180
3 1,205 2,500 2,380
4 1,205 5,000 4,460
5 1,180 10,000 8,500
6 2,330 625 345
7 2,400 1,250 735
8 2,400 2,500 1,320
9 2,355 5,000 2,380
10 2,400 10,000 4,850
11 2,650 1,250 1,135
12 1,880 2,500 1,724
13 1,570 5,000 3,611
14 2,850 5,0 00 4,154
15 1,100 10,000 9,500
16 1,600 10,000 8,440
17 1,600 10,000 5,940
47
A summary of these data and the respective
distribution coefficientsf. or equilibrium K-values are
shown in Table 6.
Table 6. Mole Fractions of Phenol in Water (x), Mole
Fractions of Phenol in Carbon Dioxide (y),
and Equilibrium K-values.
Run
Density
(g/cc)
Mole
X
Fractions
y
K-values
y/x
I .345 .000114 .00000913 .0799 .
2 .340 .000226 .0000235 .104
3 .335 .000456 .0000403 .0884
4 .335 .000854 .000181 .212
5 • .300 .00163 .000581 .357
6 .794 .0000661 .0000402 .608
7 .803 .000141 .0000732 .520
8 .805 .000253 .000167 .662
9 .797 .000456 .000375 .822
10 .803 .000929 .000732 .789
11 .817 .000217 .0000275 .1,26
13 .664 .000692 .000239 .345
14 .830 .000795 .000458 .575
15 .222 .00182 .000849 .469
16 .670 .00162 .00105 .647
17 .670 .00114 .000690 .607
Figures 7 and 8 show three-dimensional plots of that
data, viewed from 30 degrees above the pressure-composition
plane and rotated horizontally 45 degrees and 135 degrees,
respectively. These plots were generated using the SYMAP
and SYMVU library routines on the MSU CP6 computer and
drawn on a CALCOMP plotter. All seventeen data points were
used to generate each plot, showing a general tendency for
Figure 7. Three Dimensional Pressure-composition Plot of Phenol/
Water/Carbon Dioxide System at 40 Degrees Celsius, Rotated
45 Degrees.
Mole
fraction
of
phenol
in
carbon
dioxide)
3000
2000
1000
Figure 8. Three Dimensional Pressure-composition Plot of
Phenol/Water/Carbon Dioxide System at 40 Degrees
Celsius, Rotated 135 Degrees.
v:
x
HOO'
1 0 4 4 8 0
904 . 82
754 . 02
603-22
452 41
301 61
I 50 -80
0 00
Mole
fraction
of
phenol
in
carbon
dioxide)
50
a large increase in y (moles phenol per mole supercritical
carbon dioxide) with increasing x (moles phenol per mole
water) and a similar tendency for increasing y with
increasing pressure. Since the solubility of phenol in
supercritical carbon dioxide is mainly a function of
density it is expected that the increase in y, at a
constant x should level off at very high pressures, with
most of the increase taking place between 1000 and 2000
psi. The unexpected waviness of the 3-D surface is most
likely an indication of inherent experimental error.
Figure 9. shows the equilibrium data measured near
1200 and 2400 psi. The correlation for each curve was
determined using the quadratic form of the Method of Least
Squares, as described by Neter and Waterman (40), are:
At 1200 psi, y = 2.44*10-5 - 0.0883x + 285.6x2
At 2400 psi, y = -3.73*10-5 + 0'.9007x - 74.72x2
These equations were then used to calculate values
for y at even intervals of x at 1200 and 2400 psi,
corresponding to carbon dioxide densities of 0.33 and
0.803 g/cc, respectively. Use of the equations for values
of x above 0.0016 was avoided because of the apparent
discrepancy in the shapes of the two curves in that
region, and also because it served no useful purpose in
-(
Mo
le
Fr
ac
ti
on
of
P
he
no
l
in
C
ar
bo
n
Di
ox
id
e)
51
For 6 experimental points near 1200 psi
y = 2.44 X IO-5- 0.0883x + 285.6x2
0.0012 I
For 5 experimental points/near 2400 psi
-3.73 X 10 + 0.9007x - 74.72x
Range of extraction
run experiments0.0010 -
0.0008 -
0.0006 -
0.0004 -
0.0002 -
0.0005 0.0010 0.0015 0.0020
x ( Mole Fraction of Phenol in Water)
Figure 9. Phenol/Water/Carbon Dioxide Equilibria
Measured Near 1200 and 2400 psi, at
40 Degrees Celsius.
52
later analyses of the extraction run data. The values are
shown in Table 7 and plotted in Figure 10. The assumption
of a linear curve on semi-log paper, between the densities
of 0.33 (1200 psi) and 0.803 (2400 psi) was based on data
obtained from Van Leer and Paulaitis (31). Their data,
also plotted in Figure 10, shows the equilibrium
concentrations of phenol in supercritical carbon dioxide
at 360C above pure, solid phenol, resulting in a nearly
linear curve for densities ranging from 0.84 to below
0.40 g/cc. Although the temperature is slightly lower than
the 40°C used throughout this research., both temperatures
are below the melting point of phenol, and the published
data should give an excellent idea of the ultimate
solubility of phenol in supercritical carbon dioxide.
Table 7. Calculated Values
Phenol in Carbon
and 0.8 03. g/cc.
of y (Mole Fraction of
Dioxide) at 0.33 g/cc
X Y at 0.33 g/cc y at 0.803 g/cc
0.0002 . 0.0000182 0.000140
0.0004 0.0000348 0.000311
0.0006 0.0000742 0.000476
0.0008 0.000137 0.000635
0.0010 0.000222 0.000789
0.0012 0.000330 0.000936
0.0014 0.000461 0.00108
0.0016 0.000614 0.00121
(M
ol
e
Fr
ac
ti
on
of
P
he
no
l
in
C
ar
bo
n
Di
ox
id
e)
53
<00 Pure
Data taken from Van Leer
and Paulaitis (31)
.005-
. 002-
.001-
.0005-
. 0002-
. 0001-
.00005-
.00002-
.00001
Carbon Dioxide Density (g/cc)
Figure 10. Calculated Values of x (Mole Fraction of Phenol
in Water) and y (Mole Fraction of Phenol in
Carbon Dioxide) for the Phenol/Water/Carbon
Dioxide System, for Varying Densities, at
40 Degrees Celsius.
54
Points were taken from Figure 10 and reconstructed in
Figure 11 to show the relationships of y versus x at
varying densities. This family of curves is shown to get
a perspective of the data from another dimension.
Figure 12 shows a family of curves based on the calculated
distribution coefficient, K, at varying values of carbon
dioxide density and x. The curves show that K tends to
converge near a value of 0.8 as the CC^ density
approaches 0.8, at the experimental conditions of 40°C
and 2400 psi. The curves shown in Figures 10, 11, and 12
should prove useful in predicting extraction capabilities
for counter-current, co-current, and mixer-settler
processes.
Counter-current Extraction
Equilibrium data gathered near 1200 psi provide an
opportunity for analyzing the extraction data gathered in
the counter-current run. The operating conditions and flow
schematic are shown in Figure 13.
By mass balance yg was calculated to be 0.000704
gmole phenol/gmole COg. The operating line obtained
from this data and the equilibrium curve for the '1200 psig
data are shown in Figure 14. The equilibrium curve was ,
determined using the six equilibrium data points nearest
1200 psi, including Runs numbered I, 2, 3, 4, 5, and 15.
/
(M
ol
e
Fr
ac
ti
on
of
Ph
en
ol
in
Ca
rb
on
Di
ox
id
e)
55
0.002 CO^ density = 0.803 g/cc
Region consisting of pressures
higher than 2400 psi. "0.001
0.0005 -
0.0002
0.0001 -
CO9 density = 0.330
0.00005 - Region consisting of
pressures lower than
1200 psi. Both carbon
dioxide densities and y
are extremely non-linear
in this region.0.00002 -
0.00001
0.0004 0.0008 0.0012 0.0016
x (Mole Fraction of Phenol in Water)
Figure 11. Graphically Determined Values of Carbon
Dioxide Phase Density and Composition for
the Phenol/Water/Carbon Dioxide System, for
Varying Values of x (Mole Fraction of Phenol
in Water), at 40 Degrees Celsius.
Ca
lc
ul
at
ed
Di
st
ri
bu
ti
on
Co
ef
fi
ci
en
t
Mole fraction of phenol in carbon dioxide
Mole fraction of phenol in water
y/x
0.4 -
x=0.0002
0.2 -
Carbon Dioxide Density (g/cc)
Figure 12. Calculated Distribution Coefficients for Varying Values of Carbon
Dioxide Density, at 40 Degrees Celsius.
57
A
y2 = 0.000704 x2 = 0.00191
T = 40° C
P = 1130 psig
G = 13.38 gmole/hr
L = 7.83 gmole/hr
V 1 = o.o X1 = 0.000709
> f
X1 - Mole fraction of phenol in water leaving the
extraction column (mole/mole)
x2 - Mole fraction of phenol in water entering the
extraction column (mole/mole)
Y1 - Mole fraction of phenol in carbon dioxide entering
the extraction column (mole/mole)
y2 - Mole fraction of phenol in carbon dioxide leaving
the extraction column (mole/mole)
T - System temperature (degrees Celsius)
P - System pressure (psig)
G - Molar flow rate of carbon dioxide (gmole/hr)
L - Molar flow rate of water (gmole/hr)
Figure 13. Flow Schematic for Counter-current Extraction
Run.
(M
ol
e
Fr
ac
ti
on
of
Ph
en
ol
in
Ca
rb
on
Di
ox
id
e)
58
0.0010 y* - Gas-phase mole fraction corresponding
to equilibrium with the bulk liquid
composition.
0.0008 - T
0.0006
0.0004 -
0.0002 -
40° C
1130 psig Equilibrium curve y/ (Approximate)—
Corresponding
y.*
0.0005 0.0010 0.0015 0.0020
x (Mole Fraction of Phenol in Water)
Figure 14. Working Curves for Obtaining Gas-phase
Composition Values.
59
Since the equilibrium curve is nearly linear, one of the
simplest means of calculating the number of overall gas
phase transfer units, Nq ,^ is to use a graphical
integration technique. The rationale for using overall
transfer numbers, arid their derivation is explained in
detail by Sherwood, et al.’(41), and is summarized below.
Since such column characteristics as local gas and
liquid phase mass transfer coefficients, fluid dynamics,
and wetted interfacial areas are generally not known, and
specifically unknown in this case, the analysis of the
column is better accomplished by evaluating overall
characteristics. By using hypothetical interfacial,
compositions the overall gas-phase coefficient can be
related to the mass flux, the pressure, (P) and the
difference betweeri y and y* :
. N = K P(y*-y),
where y is a point on the operating line, and y* is the
gas-phase mole fraction corresponding to equilibuium with
the bulk liquid composition, x, as shown in Figure 14. The
mass flux is N, and K is the overall gas-phase coefficient.
In their analysis, Sherwood, et al. (41) also show that the
differential section of a column is determined by:
G dy
dh = - -------- . ,
N a(l-y)
where G is the molar flow rate of the gas phase, and the
60
constant, a, is the effective mass-transfer surface area .
per unit of packed volume.
By combining the two equations:
h
*2
K
yI
G dy
P a(l-y) (y'*-y)
r
where h is the total height of a column. Multiplying and
dividing by Y*BM gives:
y*BM ay
(1-y)(y*-y)
9
where y*BM
(l-y)-(l-y*)
ln[(l-y)/(l-y*) ]
The left-hand term under the integral defines the
overall gas-phase height of a transfer unit, Hq ,^ and
the right-hand term defines the number of overall gas-phase
transfer units, :
y*BMdy
(1-y)(y*-y)
If an average value of H may be used then H3 og og
can be calculated from the known column height and the
evaluated . Also, if two runs had been possible the
product, K a, could then be calculated by simultaneous
solution.
61
The integral representing. Nq^ can be evaluated
graphically by plotting
BM versus
(1-y)(y*-y)
and calculating the area under the curve. For dilute
solutions, as in this case, Y*gM is approximately
equal to 1.0 and it will be treated as such here. The
values of y and y* are obtained from the enlarged graph
of the operating line and equilibrium curve shown in
Figure 14.
Table 8 shows the data obtained from Figure 14, and
plotted in Figure 15. The area under the curve in
Table 8. Values for
Gas-phase
Calculating the Number of Overall
Transfer Units.
Y (1-y) Y* (Y * —Y) 1/t(1-y)(y*-y)]
0.00000 1.00000 .000120 .000120 8,333
0.00005 0.99995 .000155 .000105 9,524
0.00010 0.99990 .000198 .000098 10,205
0.00015 0.99985 .000238 .000088 11,365
0.00020 0.99980 .000282 .000082 12,198
0.00025 0.99975 .000330 .0000080 12,503
0.00030 0.99970 .000378 .000078 12,824
0.00035 ' 0.99965 .000430 .000080 12,504
0.00040 0.99960 .00048.2 .000082 12,200
0.00045 0.99955 .000540 .000090 11,116
0.00050 0.99950 .000602 .000102 9,809
0.00055 0.99945 .000665 .000115 8,700
0.00060 0.99940 .000736 .000136 7,357
0.00065 0.99935 .000805 .000155 6,456
0.000704 0.999296 .000890 .000186 5,380
In
te
gr
at
io
n
fa
ct
or
62
14,000
12,000 -
10,000 -
8,000 -
Area under curve = N
Number of overall
gas-phase transfer units
6,000 -
7.25
* - Gas-phase mole fraction
corresponding to equilibrium
with the bulk liquid
composition.
4,000-
2,000 - ln[(1-y)/(1-y*)]
0.0004 0.0006 0.00080.0002
y (Mole fraction of phenol in carbon dioxide)
Figure 15. Graphical Integration of the Number of Overall
Gas-phase Transfer Units.
63
Figure 15 is N and is calculated to be 7.25. This°g
indicates that the column contributes the equivalent of
7.25 equilibrium stages when operated at these specific,
conditions where the equilibrium line is approximately
linear. Since the column height is 12 inches, the overall
height of a gas-phase transfer unit is 1.66 inches. This
indicates that the column provides excellent exchange
capabilities when operating under, counter-current
conditions.
Phenol/Water Co-current Extraction
The twenty-two successful co-current extraction runs,
including both phenol/water standards and wastewater,
formed the bulk of the extraction data gathered during this
research. These data were reduced to a common form, shown
in Table 9. The seventeen extraction experiments using
phenol/water standards were run to determine the effects of
column packing, residence time, and system pressure on the
extraction efficiency. The extraction efficiency, in this
case, is defined as one hundred times the ratio of the
distribution coefficient measured at the extraction
conditions, K^, divided by Kt, the distribution
coefficient as determined by the equilibrium experiments.
The extraction run values of y (the mole fraction of phenol
in the carbon dioxide phase) were calculated using a mass
Tablei 9. Summary of Co-current Extraction Run Data
Run Average Average Net C00/H0(D Total Final Percent
No. Pressure Specific Flow Cone. of
(psig) Gravity g/g mol/mol .ml/ml ml/min mg/1 Original
PhenOl/Water Standard with 1/4 inch Raschig Rings
I 1935 0.740 4.65 1.90 6.28 4.08 4370 43.7
2 2425 0.806 1.75 0.71 2.17 1.69 5700 57.0
3 2380 0.800 5.08 2.08 6.34 3.89 3850 38.5
4 2357 0.797 11.8 4.83 14.8 7.97 2420 24.2
5 2825 0.831 2.65 1.08 3.14 1.83 4170 42.9
6 2800 0.828 6.07 2.49 7.22 3.63 3600 36.0
Phenol/Water Standard with 1/16 inch Glass Helices
I 1497 0.642 4.63 1.89 7.21 4.73 45 50 45.5
8 1517 0.647 2.9 2 1.19 4.51 3.11 5390 53.9
9 1810 0.720 2.39 0.98 3.3 2 2.40 5245 52.5
10 1825 0.722 5.80 . 2.37 8.03 4.92 4240 42.4 .
11 • 2008 0.752 3.26 1.34 4.34 2.86 4840 48.4
12 2010 0.752 6.84 2.80 9.10 5.43 3905 39.1
13 2410 0.804 3.47 1.42 4.32 2.78 4720 47.2
14 2400 0.803 6.80 2.78 8.47 4.95 2950 29.5
15 2375 0.800 14.0 5.72 17.5 9.63 2250 22.5
16 2816 0.830 3.02 1.24 3.58 2.30 4340 43.4
17 2815 0.830 8.88 3.63 10.6 5.69 2520 25.2
Coal Gasification Wastewater With 1/16 Inch Glass Helices
18 2390 0.800 7.63 3.12 9.54 5.37 1200 24.0
19 2400 0.800 6.89 2.82 8.61 4.97 1410 28.2
20 2450 0.802 8.19 3.35 10.2 5.27 1170 23.4
21 2470 0.804 2.89 1.18 3.59 . 2.26 2005 40.1
22 2425 0.801 21.13 8.65 .26.4 14.3 980 19.6
65
balance based on a known x (the mole fraction of phenol in
the water phase) and the relative net flow rates of carbon
dioxide and phenol/water solution. An experimental value
for K (K= y/x) was then calculated for each run.
To determine the theoretical K value it was necessary
to use a trial and error approach based on an initial guess
of an x value. A value for y was then calculated, again
based on a mass balance. Then, by referring to Figure 10,
the theoretical value of y was determined for the guessed
value of x. A new value for x was then estimated, if
necessary, and the process repeated until the calculated
and plotted value of y were essentially equal. Respective
K values were then calculated for each extraction run.
Theoretical and actual x, y data and K values are listed in
Table 10, and examples of the calculation procedures are
shown in Appendix B.
Upon analyzing the data it became apparent that some
of the calculated K values, obtained during runs with
relatively long residence times, were actually higher than
the theoretical K values. Normally, the highest possible
extraction efficiency for a co-current flow column would be
100%, when the two phases would be exactly in equilibrium.
This indicated that there was probably some significant
degree of error involved in the equilibrium experiments
resulting in low K values. It was felt, however, that the
Table 10. Converted but as the phenol concentration
decreases in the exiting water the equilibrium
concentration in the carbon dioxide phase also decreases.
This is an expected result based on the equilibrium
experiments.
All 17 of the phenol/water extraction data points are
shown in Figure 20, with the two types of packing identi
fied. A plotted curve represents the Method of Least
Squares equation of best fit for the entire set of points,
and has an r-value of 0.77. Individual equations of best
fit were calculated for the helices and the Raschig Rings.
These equations are listed and plotted in Figure 20.
Again, the type of packing appears to have little, if any,
influence on the extraction efficiency at these .
Ph
en
ol
Re
mo
va
l
(m
g
Ph
en
ol
/g
Ca
rb
on
Di
ox
id
e)
74
2 . OOi
(Run #16)
1.75-
Relative mass flow rate of
approximately 3.0 g CO9Zg water.
(Run #8)
1.50-
1.25-
(Run #14)
1 . 00-
(Run #12)
Relative mass flow rate of approximately
6.8 g CO9Zg water.
0.75-
0.50
1500 1750 2000 2250
Pressure (psig)
• 2500
Figure 19. Phenol Extraction Capabilities of Carbon Dioxide
with Respect to Pressure, at 40 Degrees Celsius.
75
Al I points*:
Er= 50.24 tR + 50.11
(□ )- Experiments using glass
helices:
55.01 t
( O ) - Experiments using
Raschig Rings:
43.31
Points represent all 17 experiments using
phenol/water standards, at system pressures
varying from 1500 to 2800 psig.
0 0.20 0.40 0.60 0.80 1.00
tR (Residence Time Relative to Run #2)
Figure 20. Summary of Normalized Extraction Efficiencies
and Residence Times for All Co-current
Experiments Using Phenol/Water Standards.
76
experimental conditions. However, increasing the
residence time allows the system to more closely approach
equilibrium conditions.
Equations of best fit were plotted for data obtained
using both types of packing at 2400 and 2800 p.si. These
curves and their respective r-values are plotted in Figure
21. In addition, an overall equation of best fit for all
10 points is listed and plotted. With an r-value of 0.88,
the overall equation has almost as good a fit as the
individual curves. It is possible that increasing the
system pressure from 2400 to 2800 psi. has only marginal
benefit since the SCF density increase is only about three
percent. The expected increase in solvent power could be
hidden by relatively small experimental error.
Figure 22 shows all 17 data points, undifferentiated
with respect to packing or pressure. The same data points
are shown in Figure 23, differentiated by experimental
system pressure. The two figures show the overall
dependence of the extraction efficiency upon residence
time. For both types of packing and at all of the
experimental test pressures an increase in residence time
increased the extraction efficiency. Figures 22 and 23
indicate that varying the system pressure or the column
packing has an insignificant affect on the ability of the
system to approach equilibrium, at these test conditions.
77
All four experiments run at 2800 psig:
En= 67.34 t r = 0.92
Equation of
^-best fit for
all 10 points:
D Er= 47.66 tR + 47.36
r = 0.88
All six experiments run at 2400 psig:
Er= 38.75 tR + 51.78
r = 0.90
A - 2400
□ - 2400
O - 2 800
O - 2800
0 0.20 0.40 0.60 0.80 1.00
tR (Residence Time Relative to Run #2)
Figure 21. Equations of Best Fit for Normalized Extraction
Efficiencies and Residence Times for Co-current
Phenol/Water Extraction Experiments, at 2400
and 2800 psi, and 40 Degrees Celsius.
(P
er
ce
nt
Ef
fi
ci
en
cy
Re
la
ti
ve
to
Ru
n
#5
)
78
100
80 -
60 -
50.24 t + 50.7740 -
0.77
20 -
0.20 0.40 0.60 0.80 1.00
(Residence Time Relative to Run #2)
Figure 22. Equation of Best Fit for Normalized Extraction
Efficiencies and Residence Times for All
Co-current Experiments Using Phenol/Water
Standards, at 40 Degrees Celsius.
(P
er
ce
nt
Ef
fi
ci
en
cy
Re
la
ti
ve
to
Ru
n
#5
)
79
40 -
0 - 1500 psi, Glass Helices
o - 1800 psi, Glass Helices
A — 2000 psi, Glass Helices and Raschig Rings
Q - 2400 psi, Glass Helices and Raschig Rings
□ - 2800 psi, Glass Helices and Raschig Rings
T ™
0.2()
~Y~
0.40
"I
0.60
~T~
0.80
tR (Residence Time Relative to Run #2)
Figure 23. Plot Showing Values for Normalized Extraction
Efficiencies and Residence Times for All
Co-current Experiments Using Phenol/Water
Standards, at 40 Degrees Celsius,
Differentiated by Test Pressure.
80
Wastewater Co-current Extraction
Data gathered in the extraction experiments using coal
gasification wastewater were reduced and normalized in the
same manner as the data relating to the phenol/water
standards and are shown in Tables 9 and 10. Note that the
wastewater extraction efficiencies are relative to the
standards, and that one of the five runs has an efficiency'
greater than 100%. Four of the five runs have efficiencies
greater than or equal to 90% indicating that there are
possibly some components in the wastewater that contribute
to the ability of carbon dioxide to extract phenol. Figure
24 shows a plot of the five wastewater extraction effi
ciency data points. Several types of equations were tried
to determine the type of curve that would best describe the
experimental data. A linearized version of a non-linear
reciprocal equation, proposed by Heller (42), was chosen.to
obtain a reasonably close fit of the points. This equation
has the form:
y = x/(a + bx).
This reciprocal equation was linearized, and solved with the
least squares program, by taking y equal to 1/y and x equal
to 1/x and solving for the equation:
(1/y) = b + (a/x).
The resulting equation, having an r-Value, of 0.98, is ■
81
5 80
o 40 tD/(0.001542 + 0.006068 t_)
0.98
^ 20
0.800.20 0.60
(Residence Time Relative to Run #2)
Figure 24. Equation of Best Fit for Normalized Extraction
Efficiencies and Residence Times for
Experiments Using Coal Gasification Wastewater,
at 40 Degrees Celsius.
82
plotted in Figure 24, along with a listing of the actual
equation. Other non-linear equation forms that were tried,
but discarded, included a. simple quadratic (which yielded a
concave downward parabola) and an exponential curve of the
form, y = a(xD) (having an r-value of 0.92).
Finally, Figure 25 shows all 22 co-current extraction
data points, differentiated by points representing phenol/
water standards and wastewater experiments. The wastewater
points show clearly higher extraction efficiencies,
relative to the standards.
(P
er
ce
nt
Ef
fi
ci
en
cy
Re
la
ti
ve
to
Ru
n
#5
)
83
120 -i
□
40 -
20
□ - Coal Gasification Wastewater Experiments
O - Phenol/Water Standard Experiments
CdW
0 0.20 0.40 0.60 0.80 1.00
tR (Residence Time Relative to Run #2)
Figure 25. Plot Showing Values for Normalized Extraction
Efficiencies and Residence Times for All
Co-current Experiments, at 40 Degrees Celsius,
Differentiated by Phenol/Water Standards and
Coal Gasification Wastewater.
84
SUMMARY
The principal objective of this research was to
investigate the feasibility of using supercritical carbon
dioxide to extract phenols and phenol derivatives, from coal
gasification wastewater. Five major steps were undertaken
to achieve that goal:
1. Measurement of phase equilibria for the phenol/
water/ carbon dioxide ternary system.
2. Design, construction, and modification of the
extraction equipment.
3. Measurement of column effectiveness using counter-
current extraction techniques.
4. Measurement of extraction efficiencies using
co-current extraction of phenol from phenol/water
standards.
5. Measurement of extraction efficiencies using
co-current extraction phenol and phenol derivatives
from coal gasification wastewater.
Seventeen phase equilibrium experiments were run to
obtain the basic information needed to analyze the
effectiveness of operating the extraction column at various
flow rates and pressures. Three-dimensional plots of the
data were generated to provide visual perspectives of the
effects of varying system pressure and composition. The
data show the expected general tendency for increased
85
solubility of phenol in carbon dioxide as the pressure is
increased and as the concentration of phenol in the water
phase is increased.
The equilibrium data is unfortunately subject to a
number of possible errors due to the inability to directly
measure the concentration of phenol in the carbon dioxide
phase. However, it is believed that the results are
reasonably indicative of actual values and that these data
can be used only for general interpretation of extraction
efficiencies.
The successful operation of the extraction .equipment
was preceded by an extensive period of equipment design,
construction, trial runs, and modifications. A number of
temperature and pressure controls were needed to properly
maintain the desired system parameters during the course of
an extraction run. Water and carbon dioxide flow rates were
measured as accurately as possible and all systems were
designed and built to provide for operator safety at
pressures up to 5000 psi. Major obstacles requiring
equipment design modifications included carbon dioxide flow
control and measurement, sample collection, extraction
column hydraulics, and proper heating of the carbon dioxide/
phenol stream as it was depressurized to atmospheric
pressure.
86
Approximately 30 attempts were made to counter-
currently extract phenol from a phenol/water standard. Many
of the equipment modifications mentioned above were required
to achieve the single successful experiment at 1130' psig and
40°C . Counter-current runs attempted at higher
pressures resulted in severe column flooding due to
relatively poor carbon dioxide flow control and very narrow
density differentials. Although limited in number, the one
counter-current extraction experiment shows that
supercritical carbon dioxide can be very effective in
removing phenol from water, at easily obtained pressures,
using commonly available industrial equipment.
Seventeen successful co-current extraction experiments
were run at pressures varying from 1500 to 2800 psi, using
two types of packing. Each experiment's data set was
analyzed to determine the approach to theoretical
equilibrium, or the relative efficiency. The type of
packing had little or no effect on column effectiveness.
However, the amount of residence time available for mass
transfer significantly affected relative efficiencies.
Five co-current extraction experiments were run using
coal gasification wastewater. Phenol extraction was
measured as a percentage of the original phenol concen
tration in the wastewater. Extraction efficiencies were
calculated based on the approach to theoretical equilibrium
87
as determined by the phenol/water/carbon dioxide phase
equilibrium data gathered as part of this research. Four of
the five runs indicated that extraction efficiencies of 90%
or greater had been achieved. One of those four was
greater than 100%, indicating that there could be some
component in the wastewater that enhances the solubility of
phenol in supercritical carbon dioxide, or that there is
additional experimental error due to unknown phenol
concentrations in beginning and final water samples.
88
'RECOMMENDATIONS FOR FUTURE STUDY
There is a great deal of potential for improving the
steady-state counter-current extraction process.
Additional effort should be made to provide a workable
column that can operate dependably between 1000 and 1500
psi. This column should have automatic temperature
control, precise feed volume control of both carbon dioxide
and wastewater, proper feed distribution, and level control
for exiting liquids. A packed column, approximately two to
four inches in diameter and ten to twenty feet tall would
provide a safe, relatively inexpensive means of
investigating supercritical continuous counter-current
extraction. The liquid level should be controlled by a
dump valve operation, either automatic or manual, but based
on an observed level. A sight glass suitable for the
desired pressure range, or a level indicator could be used.
Liquid samples can then be taken during the entire course
of the run via the dump line. Column internals could be
obtained as standard industrial equipment.
Ongoing research in the area of supercritical fluid
extraction (SCFE) of phenols from wastewater should attempt
to measure concentrations of extracted materials in both
89
the SCF phase and the liquid phase. This provides a means
of cross-checking measurements using material balances and
is inherently more accurate. Concentrations in the SCF
phase can be measured using thin layer chromatography (43)
or supercritical fluid chromatography (44). Either method
would require a direct hookup to a slip stream of the SCF
exiting the column.
The ability to analyze the SCF phase would be par
ticularly helpful during phase equilibrium investigations.
Continued research in this area is necessary to refine the
data already gathered, and to properly analyze extraction
column effectiveness.
Additional research in co-current extraction should
seriously consider a mixer-settler approach. This should
reduce the time required for the SCF and liquid phases to
reach equilibrium, and eliminate the unknowns of column
hydraulics and distribution. Proper control of feed rates
and feed temperatures could make this method of extraction
a convenient tool for obtaining equilibrium data,
especially if a method of recycle were built in.
The effects of varying system temperatures should be
investigated. Generally, as system temperatures are ,
increased, a substantial increase in pressure is required
to achieve what is believed to be desirable carbon dioxide
densities. On the other hand, as SCF densities approach
90
normal liquid densities, the increases in temperature tend
to increase solvent power, mimicking the characteristics of
liquid/liquid extraction. However, little is known about
the actual effects of temperature for this particular SCF/
liquid system arid an optimum temperature should be
determined.
Finally, further research should investigate the
ability of supercritical carbon dioxide to extract other
major components of coal gasification wastewater. These
include ammonia, COD (of which phenol is a constituent),
hydantoins (45), ketones, aldehydes, and cyanide. Proof of
these extraction capabilities could well make supercritical
fluid extraction a viable industrial process, especially at
the near critical conditions of carbon dioxide.
91
NOTATIONS
92
NOTATIONS
a - Effective mass transfer surface area per unit of
packed volume
BPR - Back pressure regulator
CO2 - Carbon dioxide
CP - Critical point
E - Efficiency (KA/KT times 100)
ER - Relative efficiency
FI - Flow indicator
G - Molar flow rate of carbon dioxide (gmole/hr)
GC - Gas chromatograph
- Height of an overall gas-phase transfer unit
H2O - Water
K - Distribution coefficient (y/x)
K - Overall gas-phase coefficient
- Actual measured distribution coefficient
CI - Theoretical distribution coefficient
L - Molar flow rate of water (gmole/hr)
N - Mass flux
NCL - Near critical liquid.
Nq^ - Number of overall gas-phase transfer units
P — System pressure
Pc - Critical pressure
93
PI - Pressure indicator
Pr - Reduced pressure, P/Pc
r - Correlation coefficient for Method of Least Squares
SCF - Supercritical fluid
SCFE - Supercritical fluid extraction
T - System temperature
Tc - Critical temperature
TI - Temperature indicator
Tj. - Reduced temperature, T/Tc
tR - Relative residence time
x - Mole fraction of phenol in the liquid phase
x, - Mole fraction of phenol in water leaving
counter-current extraction column
x„ - Mole fraction of phenol in water entering
counter-current extraction column
y - Mole fraction of phenol in the carbon dioxide phase
y -.Mole fraction of phenol in carbon dioxide entering
counter-current extraction column
y - Mole fraction of phenol in carbon dioxide leaving
counter-current extraction column
y* - Gas-phase mole fraction of phenol corresponding to
equilibrium with the bulk liquid composition
y*BM - Log mean average of (1-y) and (1-y*)
94
REFERENCES CITED
95
REFERENCES CITED
1. Mastertonz W. , and E. Slowinskiz Chemical Principlesz
Saunders, Philadelphia, 1977, 265-266.
2. Bott, T. R., Fundamentals of Carbon Dioxide in Solvent
Extraction, Chemistry and Industry, London, Issue
12, 1982, 394-396.
3. de Filippi, R. P., COg as a Solvent: Application to
Fats, Oils and Other Materials, Chemistry and
Industry, London, Issue 12, 1982,. 390-394.
4. Paulaitis, M. E., V. J. Kfukonis, R. T. Kurnik, and
R. C. Reid, "Supercritical Fluid Extraction",
Rev. Chem. Eng., 1982, I, 179.
5. Little, A.D. Chemical Week, 9 July 1980, 31.
6. McKugh, M., and V. J. Krukonis, "Processing with
Supercritical Fluids", AIChE Advanced Seminar,
September, 1983.
7. Scott, R. L., Thermodynamics of Critical Phenomena in
Fluid Mixtures, Ber. Bunsenges. Phys. Chem., 1972,
76, 296.
8. Scott, R. L. and P. H. van Konynenburg, State
Properties of Solutions - Van der Waals and Related
Models for Hydrocarbon Mixtures, Discuss. Faraday
Soc., 1970, 49, 87.
9. Balder, J. R. and J. M. Prausnitz, Thermodynamics of
Ternary, Liquid-Supercritical Gas Systems with
Applications for High Pressure Vapor Extraction,
Ind. Eng. Chem. Fund., 1966, 5, 449.
10. Wenzel, H. and W. Rupp, Calculation of Phase
Equilibria in Systems Containing Water and
Supercritical Components, Chem. Eng. Sci.,
1978, 33, 683.
96
11. Peng, D. Y. and D. B. Robinson, Two and Three Phase
Equilibrium Calculations for Systems Containing Water,
Can. J. Chem. Eng., 1976, 54, 595.
12. Deiters, U. and G. M. Schneider, Fluid Mixtures at
High Pressures. Computer Calculations of the Phase
Equilibria and the Critical Phenomena in Fluid Binary
Mixtures form the Redlich-Kwong Equation of State,
Berichte Bun. Gesselschaft, 1976, 12, 1316.
13. Chai, C. P., "Phase Equilibrium Behavior for Carbon
Dioxide and Heavy Hydrocarbons", Ph.D. Thesis,
University of Delaware, Newark, Delaware, 1981.
14. Prausnitz, J. M., "Phase Equilibria for Complex
Mixtures",, presented at the Third Int. conference on
Fluid Properties and Phase Equilibria for Chemical
Process Design, Callaway Gardens, Georgia, April
10-15, 1983.
15. Mathias, P. M. and T. W. Copeman, "Extension of the
Peng-Robinson Equation of State to Complex Mixtures:
Evaluation of the Various Forms of the Local
Composition Models", presented at the Third Int.
Conference of Fluid Properties and Phase Equilibria
for Chemical Process Design, Callaway Gardens,
Georgia, April 10-15, 1983.
16. Wong, J. M. and K. P. Johnston, Thermodynamic Models .
for Nonrandom and Strongly-Nonideal Liquid Mixtures,
Ind. Eng'. Chem. Fund., 1984, 23, 320-326.
17. Johnston, K. P. and C. A. Eckert, An Analytical
Carnahan-Starling van der Waals Model for Solubility
of Hydrocarbon Solids in Supercritical Fluids,
AIChE Journal, 1981, 27, 773.
18. Johnston, K. P., D. H. Zieger, and C. A. Eckert,
Solubilities of Hydrocarbon Solids in Supercritical
Fluids: The Augmented van der Waals Treatment,
Ind. Eng. Chem. Fund., 1982, 21, 191.
19. MacKay, M. E. and M. E. Paulaitis, Solid Solubilities
of Heavy Hydrocarbons in Supercritical Solvents,
Ind. Eng. Chem. Fund., 1979, 18, 149.
20. Hannay J. B. and J. Hogarth, Proc. Royal Society,
London, Ser. A, 1879, 29, 454.
97
21. Williams, D. F., Extraction with Supercritical Gases,
Review Article Number 5, Chemical Engineering
Science, Vol. 36, No. 11, pp. 1769-1788, 1981.
Printed in Great Britain, Pergamon Press Ltd. D. F.
Williams, National Coal Board, Coal Research
Establishment, Stoke Orchard, Cheltenham,
Gloucestershire, GL524RZ, England.
22. Gardner, D. S., Industrial Scale Hop Extraction with
Liquid COg, Chemistry and Industry, London, Issue
no. 12, 1982, 402-406.
23. Brogle, H., COg as a Solvent: Its Properties and
Applications, Chemistry and Industry, London, Issue
no. 12, 1982, 385-390,
24. Perry, R. H., and C. H. Chilton, Chemical Engineers'
Handbook, Fifth Edition. McGraw-Hill Co., New York,
1969, Chapter 3.
25. Vollbrecht, R., Fundamentals of Carbon Dioxide in Solvent
Extraction, Chemistry and Industry, London, Issue
. 12, 1982, 397-399.
26. Mack, H., and M. Koepsel, Parfumerie und Kosmetik,
1973, 54, 233-237.
27. Calame, J. P., and R. Steiner, Fundamentals of Carbon
Dioxide in Solvent Extraction, Chemistry and
Industry, London, Issue 12, 1982, 399-402.
28. Granlund, C. R., "Semicontinuous Supercritical Fluid
Extraction of Peat", Master of Science Thesis/ Montana
State University, 1982.
29. Smith, S. S., "Semicontinuous Supercritical Extraction",
Master of Science Thesis, Montana State University,
1983.
30. Myklebust, K. L., Jr., "Supercritical Fluid Extraction
of Peat", Master of Science Thesis, Montana
State University, 1982.
31. Van Leer, R. A. and M. E. Paulaitis, Solubilities of
Phenol and Chlorinated Phenols in Supercritical Carbon
Dioxide, J. Chem. Eng. Data, 1980, 25, 257-259.
98
32. McHugh, M. and M. E. Paulaitis, Solid Solubilities of
Naphthalene and Biphenyl in Supercritical Carbon
Dioxide, J. Chem. Eng. Data, 1980, 25, 326-329.
33. Kurnik, R. T., S. J. Holla, and R. C. Reid, Solubility
of Solids in Supercritical Carbon Dioxide and Ethylene,
J. Chem. Eng. Data, 1981, 26, 47-51.
34. Tsekhanskaya, Y. V., M. B. Iomtev, and E. V. Muskina,
Zh. Fiz. Khim.,1964, 38, 2166..
35. Kuk, M. S., and J. C. Montagna, Solubility of Oxygenated
Hydrocarbons in Supercritical Carbon Dioxide, Chemical
Engineering at Supercritical Fluid Conditions, Edited by
M. E. Paulaitis, J. M. L. Penninger, R. D. Gray, Jr.,
and P. Davidson, Ann Arbor Science Publishers, Ann Arbor
Michigan, 1983, Chapter 4.
36. Elsenbeiss, J., San Antonio, TX, Aug. 1964, Southwest
Research Institute Final Report Contract No. DA
18-108-AMC-244(A).
37. Tsekhanskaya, Y. V., M. B. Iomtev, and E. V. Muskina,
Russ. J. Phys. Chem. (Engl. Transl.), 1962, 36,
1177.
38. Paulaitis, M. E., M. A. McHugh, and C. P.-Chai, Solid
Solubilities in supercritical Fluids at Elevated
Pressures, Chemical Engineering at Supercritical Fluid
Conditions, Edited by M. E. Paulaitis, J. M. L.
Penninger, R. D. Gray, Jr., and.P. Davidson, Ann Arbor
Science Publishers, Ann Arbor Michigan, 1983, Chapter 6.
39. Francis, A. W., Ternary Systems of Liquid Carbon
Dioxide, J. Phys. Chem., 58, 1099, 1954.
40. Neter, J., and W. Wasserman, Applied Linear Statistical
Models, Richard D. Irwin, Inc., Homewood, Illinois,
1974.
41. Sherwood, T. K., R. L. Pigford, and C- R- Wilke, Mass
Transfer, McGraw-Hill Co., New York, 1975.
42. Heller, H., Choosing the Right Formula for Calculator
Curve Fitting, Chemical Engineering, February 13,
1978, 119-120.
99
43. Stahl, E., W. Schilz, E. Schutz, and E. Willing, A Quick
Method for the Microanalytical Evaluation of the
Dissolving Power Supercritical Gases, Angew. Chem. Int.
' Ed. Enlg., 1978, 17, 731-738.
44. Gere, D. R., R. Board, and D. McManigill, Supercritical
Fluid Chromatography with Small Particle Diameter Packed
Columns, Anal. Chem., 1982, 54, 736-740.
45. Olson, E. S. , J. Wi Diehl, and D. J . Miller, "Hyda.ntoins
in the Condensate Water from the Slagging Fixed Bed
Gasification of Lignite", U.S. Dept, of Energy, Grand
Forks Energy Technology Center, Grand Forks, North
Dakota, 1982.
46. Jeris, J. S. A Rapid COD Test, Water and Wastes
Engineering, 1967, 4, 89-91.
47. Standard Methods, 14th Ed., 1976, 650.
APPENDICES
101
. APPENDIX A
ANALYTICAL PROCEDURES
AMMONIA ANALYSIS USING THE ORION 901 IONALYZER
102
1. Before the analysis period starts, the instrument will
have been on and the electrode slope will have been
determined. The electrode will be in a 10 ppm solution
with added NaOH.
2. Prepare a fresh 10 ppm N standard by adding I ml of
1000 ppm standard to 98 ml of distilled water in a 150
ml beaker.
3. Prepare a blank by placing 100 ml of distilled water in
a 150 ml beaker. Place both the standard and the blank
in the tray of 23°C water. After these samples
have set for 15-20 minutes, the instrument may be
calibrated.
4. Set the instrument mode to CONCN. Set the STD VALUE
switches to 10,000. Set the SLOPE switches to the
value given the instructor (sign must be minus).
5. Rinse the electrode in distilled water by dipping it
into a beaker of distilled water two or three times.
Place the electrode in the 10 ppm standard prepared
above. Add I ml of 10 M NaOH to the solution. Use the
magnetic stirrer at all times. Press READ/CLEAR MV and
wait for a stable reading. Press SET CONCN.
6. Rinse the electrode well in distilled water.
7. Place the electrode in the blank and add I ml NaOH
stirring constantly. Wait 4 minutes and press SET
BLANK. Rinse the electrode in distilled water and
place it in the 10 ppm solution where it was
originally.
8. The instrument is now calibrated and is ready for
unknown samples. To read sample concentrations rinse
the electrode, place it in the sample, add I ml 10 M
NaOH and wait for a stable reading. Be sure to rinse
the electrode between samples. When the electrode is
not in use, it should be in the 10 ppm solution.
9. The concentration readings will be in ppm. N in the
sample in contact with the electrode.
103
10. Caution: If the READ/CLEAR MV button should be
inadvertently pushed, the calibration should be
repeated (steps 2 through 7).
11. SHUT DOWN PROCEDURE: .
a) Disconnect the electrode from the instrument and
place it in the solution labeled for overnight
storage.
b) Connect the shorting wire across the instrument
where the electrode was disconnected.
c) Set the instrument on STDBY mode.
d) Unplug the instrument.
104
CHEMICAL OXYGEN DEMAND (COD) - RAPID METHOD (46, 47)
Method
Titrimetry
Apparatus
Large hot plate preheated for one hour, thermometer
reading at least 200°C, 500 ml Erlenmeyer flasks, spoon
to hold 0.3 grams of HgSO^, glass beads.
Principle
" See "Chemical Oxygen Demand - Standard Method" (47).
In the rapid method the digesting .liquor contains both
concentrated HgSO^ and H^PO^. The sample is heated to
180OC and immediately cooled, so the test is run in about
20 minutes as compared to 3 hours for the standard method.
Interferences and Accuracy
In general, results are usually 2% to 10% lower than
the standard method. A substance which is difficult to
oxidize by the standard method (such as acetic acid,
glycine, pyridine) is less completely oxidized by the rapid
method.. For high accuracy, compare 10 replicates of the
wastewater analyzed by each method and apply a correction
factor to the rapid method.
Procedure
1. Add approximately 0.3 g of HgSO^ to a 500 ml
Erlenmeyer flask with a spoon. (With practice the
tip of a spatula can be used.) Add a few boiling
beads.
2. Pipet 10 ml or a suitable aliquot of.the sample
diluted to 10 ml into the flask (approximately 2
to 8 mg COD in sample - 400 to 1600 mg/1 - for
. best results). Mix well for 15 seconds or if
Cl is present mix I minute. IMPORTANT I
105
3. Two blanks are run by substituting 10 ml distilled
water for the sample and continuing with Steps 4 .
through 9. The average of the two titrations is
used in the calculations.
4. Carefully add 25 ml of the RAPID DICHROMATE ACID
solution to the flask and swirl the contents. Do
not use either "Dichromate Standard" nor COD
Standard Method sulfuric acid reagent.
5. Handling the flask by the neck, place the flask on
a preheated hot plate and heat to approximately
180°C. A thermometer should be immersed and
frequent swirling should be employed. The
temperature is read by tilting the flask to
immerse thethermometer sufficiently.
Approximately 5 to 8 minutes is required to reach
180°C, at which time the flask is removed from
the hot plate. With experience, 4 or 5 flasks may
be heated at one time.
6. Cautiously add approximately 300 ml of distilled
water. (The addition of water to the very hot
acid mixture can induce vigorous boiling and
spattering. Be careful.)
7. Cool the solution in a water bath to ambient
temperature.
8. Add 5 drops of ferroin indicator and titrate to
an orange end point with ferrous ammonium
sulfate, 0.05 N.
9. Two standards are run by adding 25 ml of 0.050 N-
dichromate standard, acidifying with 20 ml
concentrated sulfuric acid and continuing with
steps 5, 6, and 7.
10. Discharge into sink with copious running water and
continue to run water for at least 10 minutes
after last discharge to prevent corrosion of drain
pipes.
Calculations
The COD is calculated as follows:
(25) (0.05)
ml from Step 9
Where N equals the normality of Fe(NH^)2(SO^)2
N x 8000 x (ml from Step 3 - ml from Step 8)
COD (mg/1):
ml sample
Reagents
1. Rapid Dichromate Acid.Solution.
Dissolve 5 g K2Cr3O7 and 20 g Ag3SO4 in a solution .
consisting of I liter each of concentrated H3SO4 and
H3PO4. Label "RAPID DICHROMATE ACID SOLUTION (for
Rapid COD only)".
2. Ferrous ammonium'sulfate.
Dissolve 20 g of Fe(NH4)3(SO4)3-S H3O in distilled
water, add 20 ml concentrated H3SO4 and dilute to
I liter. This solution is standardized against
0.050 N K3Cr3O7 prior to use.
3. Ferroin indicator solution.
Dissolve 1.485 g I,10-phenanthroline monohydrate,
together with 695 mg FeSO4*'7 H3O in water and
dilute to 100 ml.
4. Mercuric sulfate.
5. Dichromate standard, 0.05 N.
Dissolve 2.4518 g K3Cr3O7 dried, primary standard
grade in distilled water and dilute to I liter.
All of above are ACS reagent grade chemicals.
APPENDIX B
SAMPLE CALCULATIONS
108
CALCULATION OF PHENOL MOLE FRACTIONS
Calculation of Mole Fraction of Phenol in Water (x).
I mg phenol I g I mole phenol 18 g water
----------- x ---- x -----;-------- x ----------- =
1000 g water 1000 mg 94 g phenol I mole water
I mg phenol -
----------- = 1.915x10 mole phenol/mole water
1000 g water
For dilute solutions (less than 1%) the mole fraction
of phenol in water is essentially equivalent to the mole ratio.
Example (Run #1):
4370 mg/1.x 1.915 x 10 ^=
0.0008369 mole phenol/mole water
109
Calculation of y (Mole Fraction of Phenol in Carbon
Dioxide) by Mass Balance.
Example (Run #2):
Let x^ = Initial mole fraction of phenol in water
Xg = Final mole fraction of phenol in water
Xg = 10000 mg/1 x 1.915 x 10 ^= 0.001915 mole/mole
Xg = 5700 mg/1 x 1.915 x 10 ^= 0.001092 mole/mole
The relative flow of carbon dioxide to water is 0.71
mole carbon dioxide to I mole water. For every mole of
water that enters and exits the column (0.001915 -0.001092)
moles of phenol are transfered to 0.71 moles of carbon
dioxide.
0.001915 - 0.001092
Therefore: y = -------------------
0.71
= 0.001160 moles phenol/mole CC^
H O
DETERMINATION OF ACTUAL DISTRIBUTION COEFFICIENT
Let CT = actual measured distribution coefficient.
mole fraction phenol in carbon dioxide phase
K = ----------------------------------------------------------------------------------------------- ------------------------------------------ =
mole fraction phenol in water phase
Example (Run #2): •
x = 0.001092 mole/mole
y = 0.001160 mole/mole (by mass balance)
CT A LFLLDDXLOLFLLDLNG A DFLXG
IDETERMINATION OF THEORETICAL DISTRIBUTION COEFFICIENT
Let K1J1 = the theoretical distribution coefficient.
Example (Data from Run #2):
Carbon dioxide density = 0.806 g/cc
Estimate x^ = 0.0012 mole/mole
Therefore y = (0.001915 - 0.0012)/0.71 = 0.00101
By referring to Figure 10, y should equal 0.00094 when
x = 0.0012.
Pick a new estimate for x^ = 0.0013
Now y .= (0.001915 - 0.0013)/0.71 = 0.000866
From Figure 10, y should now equal 0.00098.
Try x^ = 0.00122
Then y = (0.001915 - 0.00122)/0.71 = 0.000979
But Figure 10 shows y = 0.00096 for x = 0.00122
Try x^ = 0.00123
Then y = (0.001915 - 0.00123)/0.71 = 0.000965
And Figure 10 shows y = 0.00097
Since the two values of y correspond within the accuracy of
Figure 10, the theoretical x and y values are taken to be:
x = 0.00123 and y = 0.000965
and CI A LFLLLNXMOLFLLDGW A LFBbM
112
DETERMINATION OF EXTRACTION EFFICIENCY, RELATIVE
EFFICIENCY AND RELATIVE TIME.
Extraction■efficiency (E).
E A CTOCI y 100
Example (Run #2):
CT = 1.062 and CI A LFBbM
Therefore., E = KA/KT x 100
= (1.062/0.785) x 100 = 135%
Relative Efficiency (Er).
Each run's extraction efficiency was normalized
relative to the most efficient extraction run which used
the phenol/water standard. The most efficient run, Run #5,
had an extraction efficiency of 145%. For Run #2., where
E = 135%, ER = (135/145) x 100 = 93%
Relative Time.
Total column flow rates were calculated for each run
and normalized relative to the run with the slowest flow
rate (longest residence time). Run #2. For Run #2, the
combined carbon dioxide and water flow rate was 1.69
ml/min. For Run #1, the same value was 4.08 ml/min.
Therefore, the relative residence time is:
tR = 1.69/4.08 = 0.41
113
DETERMINATION OF ABSOLUTE PHENOL EXTRACTION
Example (Run #1):
Relative mass flow rate = 4.65 g carbon dioxide/g water
Initial phenol concentration in water = 10000 mg/1
Final phenol concentration in water = 4370 mg/1
10000 mg/1 - 4370 mg/1 I g water
Phenol extracted = --------=— :-------'---- x ---------
1000 g water/1 4.65 g CO^
= 1.21 mg phenol/g CO2
om)We)e sr»-rc — _
1762 10013799
Main
N378 Gartner, Ambrey Gordon
G197 Supercritical fluid
cop.2 extraction of...
DATE ISSUED TO
Main
H378
G197
cop. 2