Micropatterned Endotracheal Tubes Reduce Secretion-Related Lumen Occlusion

dc.contributor.authorMann, Ethan E.
dc.contributor.authorMagin, Chelsea M.
dc.contributor.authorMettetal, M. Ryan
dc.contributor.authorMay, Rhea M.
dc.contributor.authorHenry, MiKayla M.
dc.contributor.authorDeLoid, Heather
dc.contributor.authorPrater, Justin
dc.contributor.authorSullivan, Lauren
dc.contributor.authorThomas, John G.
dc.contributor.authorTwite, Mark D.
dc.contributor.authorParker, Albert E.
dc.contributor.authorBrennan, Anthony B.
dc.contributor.authorReddy, Shravanthi T.
dc.date.accessioned2017-06-13T16:06:32Z
dc.date.available2017-06-13T16:06:32Z
dc.date.issued2016-12
dc.description.abstractTracheal intubation disrupts physiological homeostasis of secretion production and clearance, resulting in secretion accumulation within endotracheal tubes (ETTs). Novel in vitro and in vivo models were developed to specifically recapitulate the clinical manifestations of ETT occlusion. The novel Sharklet™ micropatterned ETT was evaluated, using these models, for the ability to reduce the accumulation of both bacterial biofilm and airway mucus compared to a standard care ETT. Novel ETTs with micropattern on the inner and outer surfaces were placed adjacent to standard care ETTs in in vitro biofilm and airway patency (AP) models. The primary outcome for the biofilm model was to compare commercially-available ETTs (standard care and silver-coated) to micropatterned for quantity of biofilm accumulation. The AP model's primary outcome was to evaluate accumulation of artificial airway mucus. A 24-h ovine mechanical ventilation model evaluated the primary outcome of relative quantity of airway secretion accumulation in the ETTs tested. The secondary outcome was measuring the effect of secretion accumulation in the ETTs on airway resistance. Micropatterned ETTs significantly reduced biofilm by 71% (p = 0.016) compared to smooth ETTs. Moreover, micropatterned ETTs reduced lumen occlusion, in the AP model, as measured by cross-sectional area, in distal (85%, p = 0.005), middle (84%, p = 0.001) and proximal (81%, p = 0.002) sections compared to standard care ETTs. Micropatterned ETTs reduced the volume of secretion accumulation in a sheep model of occlusion by 61% (p < 0.001) after 24 h of mechanical ventilation. Importantly, micropatterned ETTs reduced the rise in ventilation peak inspiratory pressures over time by as much as 49% (p = 0.005) compared to standard care ETTs. Micropatterned ETTs, demonstrated here to reduce bacterial contamination and mucus occlusion, will have the capacity to limit complications occurring during mechanical ventilation and ultimately improve patient care.en_US
dc.identifier.citationMann EE, Magin CM, Mettetal MR, May RM, Henry MM, DeLoid H, Prater J, Sullivan L, Thomas JG, Twite MD, Parker AE, Brennan AB, Reddy ST. “Micropatterned Endotracheal Tubes Reduce Secretion-Related Lumen Occlusion,” Annals of Biomedical Engineering 2016 Dec;44(12):3645-3654.en_US
dc.identifier.issn0090-6964
dc.identifier.urihttps://scholarworks.montana.edu/handle/1/13049
dc.titleMicropatterned Endotracheal Tubes Reduce Secretion-Related Lumen Occlusionen_US
dc.typeArticleen_US
mus.citation.extentfirstpage3645en_US
mus.citation.extentlastpage3654en_US
mus.citation.issue12en_US
mus.citation.journaltitleAnnals of Biomedical Engineeringen_US
mus.citation.volume44en_US
mus.data.thumbpage7en_US
mus.identifier.categoryEngineering & Computer Scienceen_US
mus.identifier.doi10.1007/s10439-016-1698-zen_US
mus.relation.collegeCollege of Engineeringen_US
mus.relation.collegeCollege of Letters & Scienceen_US
mus.relation.departmentCenter for Biofilm Engineering.en_US
mus.relation.departmentChemical & Biological Engineering.en_US
mus.relation.departmentMathematical Sciences.en_US
mus.relation.researchgroupCenter for Biofilm Engineering.en_US
mus.relation.universityMontana State University - Bozemanen_US

Files

Original bundle

Now showing 1 - 1 of 1
Thumbnail Image
Name:
17-005_Micropatterned_Endotracheal_Tubes_A1b.pdf
Size:
1.02 MB
Format:
Adobe Portable Document Format
Description:
Micropatterned Endotracheal Tubes Reduce Secretion-Related Lumen Occlusion (PDF)

License bundle

Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
826 B
Format:
Item-specific license agreed upon to submission
Description:
Copyright (c) 2002-2022, LYRASIS. All rights reserved.