The accuracy of oxygen flux measurements using microelectrodes

dc.contributor.authorRasmussen, Kjetil
dc.contributor.authorLewandowski, Zbigniew
dc.date.accessioned2018-02-12T22:22:56Z
dc.date.available2018-02-12T22:22:56Z
dc.date.issued1998-12
dc.description.abstractAn electrochemical analog of a biofilm was constructed to test the accuracy of oxygen flux measurements using microelectrodes. We used a cathodically polarized graphite felt attached to the bottom of a flat plate open channel reactor as the reactive surface consuming oxygen. The oxygen flux to the felt was calculated from the polarization current. Microelectrodes were used to measure the oxygen profiles above and within the graphite felt. From the shape of the oxygen profile we evaluated the oxygen flux to the graphite felt. This provided us with two sets of data, the true oxygen flux, calculated from polarization current, and the oxygen flux estimated from microelectrode measurements. Comparing these two fluxes, for different flow velocities, showed that the fluxes evaluated from the polarization current were different from those evaluated from the oxygen profiles. The differences were likely caused by the presence of the microelectrode in the mass boundary layer and/or by the simplifying assumptions accepted in computational procedures employed to calculate oxygen fluxes. For low flow velocities, between zero and 1.0 cm s−1, the differences were velocity sensitive; the higher the flow velocity, the bigger the difference. For higher flow velocities, between 1 cm s−1 and 3 cm s−1, the flux of oxygen estimated from the microelectrode measurements was consistently approximately 80% higher than the true oxygen flux estimated from the polarization current.en_US
dc.identifier.citationRasmussen, K. and Z. Lewandowski, "The accuracy of oxygen flux measurements using microelectrodes," Water Research, 32(12):3747-3755 (1998).en_US
dc.identifier.issn0043-1354
dc.identifier.urihttps://scholarworks.montana.edu/handle/1/14362
dc.titleThe accuracy of oxygen flux measurements using microelectrodesen_US
dc.typeArticleen_US
mus.citation.extentfirstpage3747en_US
mus.citation.extentlastpage3755en_US
mus.citation.issue12en_US
mus.citation.journaltitleWater Researchen_US
mus.citation.volume32en_US
mus.data.thumbpage4en_US
mus.identifier.categoryEngineering & Computer Scienceen_US
mus.identifier.doi10.1016/s0043-1354(98)00149-3en_US
mus.relation.collegeCollege of Engineeringen_US
mus.relation.departmentCenter for Biofilm Engineering.en_US
mus.relation.departmentChemical & Biological Engineering.en_US
mus.relation.departmentChemical Engineering.en_US
mus.relation.researchgroupCenter for Biofilm Engineering.en_US
mus.relation.universityMontana State University - Bozemanen_US

Files

Original bundle

Now showing 1 - 1 of 1
Thumbnail Image
Name:
98-028_The_accuracy_of_.pdf
Size:
445.9 KB
Format:
Adobe Portable Document Format
Description:
The accuracy of oxygen flux measurements using microelectrodes (PDF)

License bundle

Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
826 B
Format:
Item-specific license agreed upon to submission
Description:
Copyright (c) 2002-2022, LYRASIS. All rights reserved.