The development and validation of a standard in vitro method to evaluate the efficacy of surface modified urinary catheters

dc.contributor.advisorChairperson, Graduate Committee: Philip S. Stewart; Darla Goeres (co-chair)en
dc.contributor.authorSummers, Jennifer Graceen
dc.date.accessioned2019-05-24T15:16:20Z
dc.date.available2019-05-24T15:16:20Z
dc.date.issued2019en
dc.description.abstractUrinary catheters are a critical medical device in modern medicine, used in almost every healthcare setting worldwide. Catheter associated urinary tract infections (CAUTI) account for 37% of all healthcare associated infections. Many surface modifications, such as antimicrobial coatings, have been proposed but none have resulted in a significant decrease in CAUTI. A variety of test methods exist to evaluate the efficacy of surface modified urinary catheters, but there is no validated in vitro standard method. This thesis reports on a standard test method which aims to replicate the two routes by which infection may occur, intraluminally and extraluminally, through two quantitative, in vitro models. The first, the Intraluminal Catheter Model (ICM), was devised to evaluate the efficacy of surface modifications to inhibit biofilm growth on the catheter lumen. The ICM was subjected to a rigorous statistical evaluation of its ruggedness. Specifically, the amount the log density and log reduction changed with small adjustments to key operational factors. Five operational factors were varied - inoculum concertation, flow of media through the catheter, pH of the artificial urine media, temperature of the incubator, and biofilm removal technique. The results of the analysis highlighted that biofilm growth is sensitive to changes in pH, which indicates that the growth media must to optimized to increase the method's ruggedness. The analysis also demonstrated that sonication was more efficient than scraping as a means to harvest biofilm from the catheter surface. With further optimization of the procedure, the ICM has potential to become a useful tool to evaluate the efficacy of surface modified urinary catheters. Three extraluminal models were tested but did not meet the statistical requirements necessary for standardization. Extraluminal infections account for 66% of all CAUTIs, therefore, the in vitro evaluation of a surface modification's ability to inhibit migration of bacteria along the extraluminal surface of a catheter is critical to fully understand how a modification will perform in the clinical setting. The development of a standardized in vitro method which reflects the physiological conditions of CAUTI will help FDA regulators to compare efficacies across products enabling them to recognize effective treatments for patients.en
dc.identifier.urihttps://scholarworks.montana.edu/handle/1/15149en
dc.language.isoenen
dc.publisherMontana State University - Bozeman, College of Engineeringen
dc.rights.holderCopyright 2018 by Jennifer Grace Summersen
dc.subject.lcshUrinary organsen
dc.subject.lcshCathetersen
dc.subject.lcshTestingen
dc.subject.lcshInfectionen
dc.titleThe development and validation of a standard in vitro method to evaluate the efficacy of surface modified urinary cathetersen
dc.typeThesisen
mus.data.thumbpage44en
thesis.degree.committeemembersMembers, Graduate Committee: Garth James; Kenneth Phillips.en
thesis.degree.departmentChemical & Biological Engineering.en
thesis.degree.genreThesisen
thesis.degree.nameMSen
thesis.format.extentfirstpage1en
thesis.format.extentlastpage139en

Files

Original bundle

Now showing 1 - 1 of 1
Thumbnail Image
Name:
SummersTheDevelopmentAndValidation2019.pdf
Size:
10.43 MB
Format:
Adobe Portable Document Format
Description:
The development and validation

License bundle

Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
826 B
Format:
Plain Text
Description:
Copyright (c) 2002-2022, LYRASIS. All rights reserved.