We describe the design and implementation of a dichroic Mueller pulsed lidar (DMPL) testbed. Operating at 532- and 1064-nm wavelengths simultaneously, the polarimetric lidar returns a full set of range-resolved Mueller matrix coefficients after a measurement period of just under 3 min. The instrument is designed as a source of reference measurements to produce proof-of-concept and high-fidelity scene models for the validation of a wide variety of other, more complex polarimetric lidar systems. Using polarimetric optimization theory and assessing the limits of our alignment and calibration, we bounded the theoretical Mueller coefficient errors to no more than 2.04%. Initial measurements show expected polarimetric features for samples characterized in the existing literature and a vast majority of measured coefficients well within the theoretical maximum error; the mean observed Mueller coefficient error across multiple scenarios was 0.86%.